A portable graphite calorimeter for onsite reference dosimetry and beam quality correction factor determination

Nicolás Gómez-Fernández, Jorge Deus-Abreu, David Maughan, Faustino Gómez, Diego M González-Castaño

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70381 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Nicolás Gómez-Fernández und 4 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Physics Laboratory, Area of Research Infrastructures, University of Santiago de Compostela, E. San Lourenzo, Santiago de Compostela, Spain. Hauptthemen laut MEDLINE: Graphite, Phantoms, Imaging, Radiometry, Calorimetry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Instituto de Salud Carlos III.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To present a portable, reliable, and fast setup graphite calorimeter for onsite absolute dose measurements at hospitals.

Methods

We have designed and built a portable graphite calorimeter optimized for fast setup and accurate measurements at clinical centers. Our system has minimum readout and auxiliary electronics and is inserted in a water fillable phantom in which chamber holders can be positioned in order to perform chamber calibrations at user's beam quality. System operation parameters were adjusted to obtain thermal equilibrium in minimum time. Our system accuracy was validated by inter‐comparison against the National Physical Laboratory (NPL) Primary Standard. Also, its portability and reliability have been tested during a measurement campaign in hospitals.

Results

Optimization of system operation parameters together with an appropriate design yielded a fully portable, fast and easy‐to‐set‐up system with only 3 h treatment room occupation time. The system exhibits thermal stability of ± 100 µK and a maximum temperature drift of 60 µK/min, allowing for precise measurements with few repetitions. Mean relative local deviation between chambers calibrated by NPL calorimeter and our calorimeter was 0.1% demonstrating our system accuracy. We measured beam quality correction factors for four chamber types and performed a successful measurement campaign in five hospitals.

Conclusions

We have developed an accurate absolute dosimeter that can be used for the determination of beam quality correction factors at clinical sites with minimal disruption of hospital workload.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

GraphitePhantoms, ImagingRadiometryCalorimetryQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedHumansRadiotherapy DosageCalibration

Stichwörter der Autor:innen

beam quality correction factorscalorimetryreference dosimetry

Substanzen

Graphite

Autor:innen und Institutionen

  • Radiation Physics Laboratory, Area of Research Infrastructures, University of Santiago de Compostela, E. San Lourenzo, Santiago de Compostela, Spain.
  • Dept de Ingeniería Mecánica, Máquinas y Motores Térmicos y Fluidos, Universidade de Vigo, Campus Marcosende, Vigo, Spain.
  • National Physical Laboratory, Teddington, UK.
  • Departamento de Física de Partículas, Universidade de Santiago de Compostela, Santiago de Compostela, Spain.
  • Fundación de Investigación Sanitaria de Santiago, Santiago de Compostela, Spain.

Förderung

  • Instituto de Salud Carlos III

Zitieren

APA

Gómez-Fernández, N., Deus-Abreu, J., Maughan, D., Gómez, F., & González-Castaño, D. M. (2025). A portable graphite calorimeter for onsite reference dosimetry and beam quality correction factor determination. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70381

Vancouver

Gómez-Fernández N, Deus-Abreu J, Maughan D, Gómez F, González-Castaño DM. A portable graphite calorimeter for onsite reference dosimetry and beam quality correction factor determination. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70381

BibTeX
@article{GmezFernndez2025,
  author = {Gómez-Fernández, Nicolás and Deus-Abreu, Jorge and Maughan, David and Gómez, Faustino and González-Castaño, Diego M},
  title = {A portable graphite calorimeter for onsite reference dosimetry and beam quality correction factor determination},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70381}
}
Kostenlos starten →