Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer

Ioannis A Tsalafoutas iD, Shady AlKhazzam iD, Mohammed Hassan Kharita

Journal of Applied Clinical Medical Physics · 2026-01 · 27 (1)

DOI: 10.1002/acm2.70331 ↗

At a glance

Journal article in Journal of Applied Clinical Medical Physics, published online on 2026-01-14. Written by Ioannis A Tsalafoutas and 2 co-authors. Institution: Medical Physics Section, OHS Department, Hamad Medical Corporation, Doha, Qatar. Major MEDLINE topics: Phantoms, Imaging, Radiographic Image Enhancement, Image Processing, Computer-Assisted. Cited 1 times so far (Crossref); 11 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

PURPOSE The exposure index (EI), the target exposure index (EI T ), and the deviation index (DI) have been defined in the IEC Standard 62494‐1 Ed.1 2008‐08. This study investigates the impact of certain acquisition parameters, the imaged anatomy, and the manufacturer's specificities on the EI of radiological images and how these may affect EI T setting procedure. METHODS Images were acquired using two digital radiography (DR) systems of two different manufacturers, using aluminum attenuators and an anthropomorphic phantom. Acquisition parameters like the tube potential (kVp), the tube loading (mAs), the exposure time, the automatic exposure control (AEC) system settings (sensor and dose level selection), the grid (with or without), the additional filtration, the field size, and the imaged anatomy were varied and their effect on the EI was quantified separately for each system. RESULTS EI is linearly related to the incident air kerma (IAK) on the detector as expected (by definition). For constant IAK, EI increases with increasing kVp. While EI in general is reduced in the presence of scatter, this may not always be the case. Under AEC operation, even the exposure time can make a difference. EI is strongly affected by the imaged anatomy in combination with the AEC sensor and field size selections, the examination protocol, and the manufacturer. CONCLUSIONS Many parameters affect the EI calculation apart from IAK. Among them, the most important are the imaged anatomy and the manufacturer. Since the EI calculation is a complex procedure, setting of the EI T values should be done with caution on a per‐examination and manufacturer basis, since the values that apply for one digital system are not always applicable to another. Furthermore, when EI is used as an image quality tool, a DI variation of at least ±2 should be allowed before a possibly meaningful red flag is activated.

Subjects

MeSH terms (MEDLINE)

Phantoms, ImagingRadiographic Image EnhancementImage Processing, Computer-AssistedHumansRadiation Dosage

Author keywords

Anthropomorphic phantomsdeviation indexdigital radiographyexposure indeximage qualitytarget exposure index

Authors and institutions

  • Medical Physics Section, OHS Department, Hamad Medical Corporation, Doha, Qatar.

ORCID profiles: Ioannis A Tsalafoutas, Shady AlKhazzam

Cite

APA

Tsalafoutas, I. A., AlKhazzam, S., & Kharita, M. H. (2026). Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer. Journal of Applied Clinical Medical Physics, 27(1). https://doi.org/10.1002/acm2.70331

Vancouver

Tsalafoutas IA, AlKhazzam S, Kharita MH. Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer. Journal of Applied Clinical Medical Physics. 2026;27(1). doi:10.1002/acm2.70331

BibTeX
@article{Tsalafoutas2026,
  author = {Tsalafoutas, Ioannis A and AlKhazzam, Shady and Kharita, Mohammed Hassan},
  title = {Exposure index in digital radiography and its dependence on acquisition parameters, anatomy, and manufacturer},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {1},
  doi = {10.1002/acm2.70331}
}
Start for free →