Together or Separate: What is best practice when imaging bilateral hands for rheumatoid arthritis examinations?

Annabel Keir iD, Vicki Braithwaite

Journal of Medical Radiation Sciences · 2023

DOI: 10.1002/jmrs.689 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Medical Radiation Sciences, 2023. Verfasst von Annabel Keir und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging, The Prince Charles Hospital, Chermside, Queensland, Australia. Hauptthemen laut MEDLINE: Radiographic Image Enhancement, Arthritis, Rheumatoid. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

A current gap identified in medical imaging (MI) literature is a standardised approach to bilateral hand examinations. This examination performed concurrently or unilaterally results in different effects on radiation dose and image quality, both of which are important to the diagnostic and follow‐up imaging of rheumatoid arthritis (RA) patients.

Methods

An experimental study was undertaken using anthropomorphic hand phantoms at the Queensland University of Technology (QUT) MI Simulation laboratory. Images of the hand were acquired individually and then concurrently with both hands together. Radiation dose was calculated by observing dose area product (DAP) reading on a digital radiography system, with the additional use of an exposure metre as a secondary data collection method. Image quality was quantified through measuring distortion caused by beam divergence, by exploring the separation of two metal rings fixed to the hand phantom.

Results

The overall radiation dose was higher for the unilateral technique by 10.15% at the digital radiography system console and 11.96% recorded by the exposure metre. In the second part of the experiment, the unilateral technique produced 0 mm of distortion when the phantom was positioned in the central part of the beam. The concurrent technique demonstrated an average of 3.65 mm of distortion, when both hands were positioned with the central part of the beam between them.

Conclusion

The unilateral technique should be performed for bilateral hand examinations. The increase in distortion from the concurrent technique is clinically significant, as the diagnostic grading of RA is measured in millimetre increments. The additional overall examination dose is minimal when compared to the improvement in image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiographic Image EnhancementArthritis, RheumatoidHumansPhantoms, ImagingQueenslandRadiation Dosage

Stichwörter der Autor:innen

Arthritisdiagnostic imaginghand bonesimagingphantomsradiation dosagerheumatoid

Autor:innen und Institutionen

  • Department of Medical Imaging, The Prince Charles Hospital, Chermside, Queensland, Australia.
  • Queensland University of Technology, Faculty of Health, Brisbane, Queensland, Australia.

ORCID-Profile: Annabel Keir

Zitieren

APA

Keir, A., & Braithwaite, V. (2023). Together or Separate: What is best practice when imaging bilateral hands for rheumatoid arthritis examinations?. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.689

Vancouver

Keir A, Braithwaite V. Together or Separate: What is best practice when imaging bilateral hands for rheumatoid arthritis examinations?. Journal of Medical Radiation Sciences. 2023. doi:10.1002/jmrs.689

BibTeX
@article{Keir2023,
  author = {Keir, Annabel and Braithwaite, Vicki},
  title = {Together or Separate: What is best practice when imaging bilateral hands for rheumatoid arthritis examinations?},
  journal = {Journal of Medical Radiation Sciences},
  year = {2023},
  doi = {10.1002/jmrs.689}
}
Kostenlos starten →