Dual‐energy CT for proton therapy: Impact of advanced slice‐wise patient‐thickness estimation methods for improved stopping‐power prediction
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70630 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Julian Schwengfelder und 3 weiteren Autor:innen. Beteiligte Einrichtung: OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine, Dresden, Saxony, Germany. Hauptthemen laut MEDLINE: Proton Therapy, Tomography, X-Ray Computed, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The direct prediction of stopping‐power ratio (SPR) from dual‐energy CT (DECT) has become gold‐standard in proton therapy. Remaining uncertainties due to patient‐size‐specific CT number variations are mitigated by calibration factors based on patient size defined as water‐equivalent thickness.
Purpose
To improve SPR prediction, two slice‐wise thickness estimation methods (TEM‐B1 and ‐B2) were compared with the previously used one (TEM‐A).
Methods
TEM‐A is using the maximum attenuation projections in x ‐ and y ‐direction, while TEM‐B1 and ‐B2 incorporate all voxels of the object to better describe non‐elliptical geometries. Simplified geometries were used to investigate TEM dependencies on several parameters (e.g., object shape, rotation). TEMs were then applied to DECT scans of cylindrical acrylic phantoms with varying diameters and to patient data. Clinical treatment plans were recalculated on generated TEM‐specific SPR datasets, and the impact of different estimated thicknesses on SPR was assessed.
Results
In contrast to TEM‐A, TEM‐B1 and ‐B2 demonstrated robustness to object shape and rotation. Couch attenuation affected all evaluated TEMs with TEM‐A being most affected. For patient scans, TEM‐B1 and B2 agreed closely but differed from TEM‐A, especially in high diameters. In obese patients, this leads to relative proton range deviations up to 0.3% when comparing TEM‐B1 and TEM‐A. In the sensitivity analysis, TEM‐B1 and ‐B2 maintained SPR uncertainties below ±3% even for cortical bone.
Conclusions
TEM‐B1 and ‐B2 reduced deviations in thickness estimation and increased robustness to object shape, overcoming the limitations of TEM‐A and improving SPR prediction accuracy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine, Dresden, Saxony, Germany.
- Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Saxony, Germany.
- Department of Radiation Oncology, University of Washington & Fred Hutch Cancer Center, Washington, Seattle, USA.
- Siemens Healthineers, Varian, Cancer Therapy Imaging, Forchheim, Bavaria, Germany.
- Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Dresden, Saxony, Germany.
- German Cancer Consortium (DKTK), Partner Site Dresden and German Cancer Research Center (DKFZ), Heidelberg, Germany.
Zitieren
APA
Schwengfelder, J., Peters, N., Wohlfahrt, P., & Richter, C. (2026). Dual‐energy CT for proton therapy: Impact of advanced slice‐wise patient‐thickness estimation methods for improved stopping‐power prediction. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70630
Vancouver
Schwengfelder J, Peters N, Wohlfahrt P, Richter C. Dual‐energy CT for proton therapy: Impact of advanced slice‐wise patient‐thickness estimation methods for improved stopping‐power prediction. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70630
BibTeX
@article{Schwengfelder2026,
author = {Schwengfelder, Julian and Peters, Nils and Wohlfahrt, Patrick and Richter, Christian},
title = {Dual‐energy CT for proton therapy: Impact of advanced slice‐wise patient‐thickness estimation methods for improved stopping‐power prediction},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70630}
}
EUVIMED