Evaluation of image quality in mobile cone‐beam CT with dose modulation using automatic exposure control: A phantom study

Keita Okazaki, Wenchao Cao, Reza Taleei, Firas Mourtada iD, Jun Li, Karen Mooney, Pramila Rani Anne, Yingxuan Chen iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70248 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Keita Okazaki und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Image Processing, Computer-Assisted. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The integration of mobile cone‐beam computed tomography (CBCT) into brachytherapy workflows offers clinical advantages such as immediate verification of applicator placement and adaptive treatment planning. These benefits require sufficient image quality to delineate applicators, target volumes, and organs at risk. A systematic evaluation of automatic exposure control (AEC) settings, radiation dose, and image quality is essential to ensure clinically acceptable imaging while minimizing patient exposure.

Purpose

This study evaluates the characteristics of AEC and its impact on image quality and radiation dose in a mobile CBCT system used for brachytherapy.

Methods

The Elekta ImagingRing CBCT system was used to scan a CatPhan phantom under two imaging protocols: Medium Dose Limit (MDL) and Ultra‐High Dose Limit (UHDL). This system employs a two‐layer mAs modulation process, consisting of preset mA values based on body mass index (BMI) and adjusted mA based on real‐time AEC. A bolus was used to simulate larger patient sizes. Real‐time x‐ray tube current at 10 degrees intervals was recorded. Image quality was evaluated using image noise, noise power spectrum (NPS), modulation transfer function (MTF), Hounsfield Unit (HU) linearity, uniformity index (UI), and contrast‐to‐noise ratio (CNR) across different protocols.

Results

AEC effectively modulated x‐ray tube current in the MDL protocol after x‐ray attenuation through the scanned phantom was measured. The UHDL protocol demonstrated greater noise reduction than the MDL. MTF values were comparable between protocols, indicating preserved spatial resolution in the MDL protocol. HU linearity was consistent across all protocols, with R 2 > 0.993.

Conclusion

AEC in mobile CBCT optimized radiation dose and image quality by adjusting tube current based on attenuation. The MDL protocol reduced radiation exposure while maintaining image quality, making it a viable option for verifying applicator placement and treatment planning in brachytherapy. The UHDL protocol achieved noise reduction with the maximum available tube current.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedBrachytherapyHumansRadiotherapy DosageRadiation Dosage

Stichwörter der Autor:innen

CBCTHDR brachytherapyautomatic exposure controlimage quality

Autor:innen und Institutionen

  • Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.

ORCID-Profile: Firas Mourtada, Yingxuan Chen

Zitieren

APA

Okazaki, K., Cao, W., Taleei, R., Mourtada, F., Li, J., Mooney, K., Anne, P. R., & Chen, Y. (2025). Evaluation of image quality in mobile cone‐beam CT with dose modulation using automatic exposure control: A phantom study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70248

Vancouver

Okazaki K, Cao W, Taleei R, Mourtada F, Li J, Mooney K, et al. Evaluation of image quality in mobile cone‐beam CT with dose modulation using automatic exposure control: A phantom study. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70248

BibTeX
@article{Okazaki2025,
  author = {Okazaki, Keita and Cao, Wenchao and Taleei, Reza and Mourtada, Firas and Li, Jun and Mooney, Karen and Anne, Pramila Rani and Chen, Yingxuan},
  title = {Evaluation of image quality in mobile cone‐beam CT with dose modulation using automatic exposure control: A phantom study},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70248}
}
Kostenlos starten →