EUVIMEDEUVIMED entdecken
Beobachtungsstudie

A prospective observational study evaluating two patient immobilisation methods in lung stereotactic radiotherapy

S Boisbouvier, I Martel-Lafay, R Tanguy, M Ayadi-Zahra

Cancer/Radiothérapie · 2024-06 · 28 (3) · 229-235

DOI: 10.1016/j.canrad.2023.08.012 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Cancer/Radiothérapie, 2024-06. Verfasst von S Boisbouvier und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Therapy Department, centre Léon-Bérard, 28, rue Laënnec, 69008 Lyon, France. Electronic address: sophie.boisbouvier@lyon.unicancer.fr. Hauptthemen laut MEDLINE: Radiosurgery, Lung Neoplasms, Immobilization, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 14 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The main objective of this study was to assess inter- and intrafraction errors for two patient immobilisation devices in the context of lung stereotactic body radiation therapy: a vacuum cushion and a simple arm support. MATERIALS AND METHODS: Twenty patients who were treated with lung stereotactic body radiation therapy in supine position with arms above their head were included in the study. Ten patients were setup in a vacuum cushion (Bluebag™, Elekta) and ten other patients with a simple arm support (Posirest™, Civco). A pretreatment four-dimensional cone-beam computed tomography and a post-treatment three-dimensional cone-beam computed tomography were acquired to compare positioning and immobilisation accuracy. Based on a rigid registration with the planning computed tomography on the spine at the target level, translational and rotational errors were reported. RESULTS: The median number of fractions per treatment was 5 (range: 3-10). Mean interfraction errors based on 112 four-dimensional cone-beam computed tomographies were similar for both setups with deviations less than or equal to 1.3mm in lateral and vertical direction and 1.2° in roll and yaw. For longitudinal translational errors, mean interfraction errors were 0.7mm with vacuum cushion and -3.9mm with arm support. Based on 111 three-dimensional cone-beam computed tomographies, mean lateral, longitudinal and vertical intrafraction errors were -0.1mm, -0.2mm and 0.0mm respectively (SD: 1.0, 1.2 and 1.0mm respectively) for the patients setup with vacuum cushion, and mean vertical, longitudinal and lateral intrafraction errors were -0.3mm, -0.7mm and 0.1mm respectively (SD: 2.3, 1.8 and 1.4mm respectively) for the patients setup with arm support. Intrafraction errors means were not statistically different between both positions but standard deviations were statistically larger with arm support. CONCLUSION: The results of our study showed similar inter and intrafraction mean deviations between both positioning but a large variability in intrafraction observed with arm support suggested a more accurate immobilization with vacuum cushion.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryLung NeoplasmsImmobilizationCone-Beam Computed TomographyPatient PositioningHumansProspective StudiesAgedFemaleMaleMiddle AgedAged, 80 and overRadiotherapy Setup ErrorsDose Fractionation, RadiationSupine PositionFour-Dimensional Computed TomographyRadiotherapy Planning, Computer-AssistedVacuum

Stichwörter der Autor:innen

ImmobilisationLungPatient positioningPositionnement patientPoumonRadiothérapie stéréotaxiqueStereotactic body radiotherapy

Autor:innen und Institutionen

  • Radiation Therapy Department, centre Léon-Bérard, 28, rue Laënnec, 69008 Lyon, France. Electronic address: sophie.boisbouvier@lyon.unicancer.fr.
  • Radiation Therapy Department, centre Léon-Bérard, 28, rue Laënnec, 69008 Lyon, France.

Zitieren

APA

Boisbouvier, S., Martel-Lafay, I., Tanguy, R., & Ayadi-Zahra, M. (2024). A prospective observational study evaluating two patient immobilisation methods in lung stereotactic radiotherapy. Cancer/Radiothérapie, 28(3), 229-235. https://doi.org/10.1016/j.canrad.2023.08.012

Vancouver

Boisbouvier S, Martel-Lafay I, Tanguy R, Ayadi-Zahra M. A prospective observational study evaluating two patient immobilisation methods in lung stereotactic radiotherapy. Cancer/Radiothérapie. 2024;28(3):229-235. doi:10.1016/j.canrad.2023.08.012

BibTeX
@article{Boisbouvier2024,
  author = {Boisbouvier, S and Martel-Lafay, I and Tanguy, R and Ayadi-Zahra, M},
  title = {A prospective observational study evaluating two patient immobilisation methods in lung stereotactic radiotherapy},
  journal = {Cancer/Radiothérapie},
  year = {2024},
  volume = {28},
  number = {3},
  pages = {229--235},
  doi = {10.1016/j.canrad.2023.08.012}
}