Evaluation of a Metal Artifact Reduction Algorithm for Image Reconstruction on a Novel CBCT Platform

Abby Yashayaeva, Robert Lee MacDonald, James Robar, Amanda Cherpak

Journal of Applied Clinical Medical Physics · 2024-11 · 25 (11)

DOI: 10.1002/acm2.14516 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 17.09.2024. Verfasst von Abby Yashayaeva und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Atmospheric Sciences, Dalhousie University, Halifax, Canada. Hauptthemen laut MEDLINE: Artifacts, Phantoms, Imaging, Algorithms, Metals. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The presence of metal implants can produce artifacts and distort Hounsfield units (HU) in patient computed tomography (CT) images. The purpose of this work was to characterize a novel metal artifact reduction (MAR) algorithm for reconstruction of CBCT images obtained by the HyperSight imaging system.

Methods

Three tissue‐equivalent phantoms were fitted with materials commonly used in medical applications. The first consisted of a variety of metal samples centered within a solid water block, the second was an Advanced Electron Density phantom with metal rods, and the third consisted of hip prostheses positioned within a water tank. CBCT images of all phantoms were acquired and reconstructed using the MAR and iCBCT Acuros algorithms on the HyperSight system. The signal‐to‐noise ratio (SNR), artifact index (AI), structural similarity index measure (SSIM), peak signal‐to‐noise ratio (PSNR), and mean‐square error (MSE) were computed to assess the image quality in comparison to artifact‐free reference images. The mean HU at various VOI positions around the cavity was calculated to evaluate the artifact dependence on distance and angle from the center of the cavity. The artifact volume of the phantom (excluding the cavity) was estimated by summing the volume of all voxels with HU values outside the 5th and 95th percentiles of the phantom CBCT with no artifact.

Results

The SNR, AI, SSIM, PSNR, and MSE metrics demonstrated significantly higher similarity to baseline when using MAR compared to iCBCT Acuros for all high‐density materials, except for aluminum. Mean HU returned to expected solid water background at a shorter distance from metal sample in the MAR images, and the standard deviation remained lower for the MAR images at all distances from the insert. The artifact volume decreased using the novel MAR algorithm for all metal samples excluding aluminum ( p < 0.001) and all hip prostheses ( p < 0.05).

Conclusion

Varian's HyperSight MAR reconstruction algorithm shows a reduction in metal artifact metrics, motivating the use of MAR reconstruction for patients with metal implants.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsPhantoms, ImagingAlgorithmsMetalsCone-Beam Computed TomographyImage Processing, Computer-AssistedSignal-To-Noise RatioProstheses and ImplantsHumansRadiographic Image Interpretation, Computer-AssistedHip Prosthesis

Stichwörter der Autor:innen

cone beam CTimage qualitymetal artifact reductionquantitative image analysis

Substanzen

Metals

Autor:innen und Institutionen

  • Department of Physics and Atmospheric Sciences, Dalhousie University, Halifax, Canada.
  • Department of Radiation Oncology, Dalhousie University, Halifax, Canada.
  • Nova Scotia Health, Halifax, Canada.

Zitieren

APA

Yashayaeva, A., MacDonald, R. L., Robar, J., & Cherpak, A. (2024). Evaluation of a Metal Artifact Reduction Algorithm for Image Reconstruction on a Novel CBCT Platform. Journal of Applied Clinical Medical Physics, 25(11). https://doi.org/10.1002/acm2.14516

Vancouver

Yashayaeva A, MacDonald RL, Robar J, Cherpak A. Evaluation of a Metal Artifact Reduction Algorithm for Image Reconstruction on a Novel CBCT Platform. Journal of Applied Clinical Medical Physics. 2024;25(11). doi:10.1002/acm2.14516

BibTeX
@article{Yashayaeva2024,
  author = {Yashayaeva, Abby and MacDonald, Robert Lee and Robar, James and Cherpak, Amanda},
  title = {Evaluation of a Metal Artifact Reduction Algorithm for Image Reconstruction on a Novel CBCT Platform},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {11},
  doi = {10.1002/acm2.14516}
}
Kostenlos starten →