EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

What is the optimal isodose line for stereotactic radiotherapy for single brain metastases using HyperArc?

Tomohiro Sagawa iD, Toshiki Ikawa iD, Shingo Ohira, Naoyuki Kanayama iD, Yoshihiro Ueda, Shoki Inui iD, Masayoshi Miyazaki, Koji Konishi

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14408 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Tomohiro Sagawa und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan. Hauptthemen laut MEDLINE: Brain Neoplasms, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiosurgery. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The study aimed to investigate the optimal isodose line (IDL) in linear accelerator‐based stereotactic radiotherapy for single brain metastasis, using HyperArc. We compared the dosimetric parameters for target and normal brain tissue among six plans with different IDLs.

Methods

This study included 30 patients with single brain metastasis. We retrospectively generated six plans for each tumor with different IDLs (80%, 70%, 60%, 50%, 40%, and 33%) using HyperArc. All treatment plans were normalized to the prescription dose of 35 Gy in five fractions which was covered by 95% of the planning target volume (PTV), defined by adding a 1.0 mm margin to the gross tumor volume (GTV). The dosimetric parameters were compared among the six plans.

Results

For GTV > 0.1 cm 3 , the ratio of brain–GTV volumes receiving 25 Gy to PTV (V 25Gy /PTV) was significantly lower at IDL 40%–70% than at IDL 80% and 33% ( p < 0.01, retrospectively). For GTV < 0.1 cm 3 , V 25Gy /PTV decreased continuously as IDL decreased. The values of D 99% and D 80% for GTV increased with decreasing IDL. An IDL of 50% or less was required to achieve D 99% of greater than 43 Gy and D 80% of greater than 50 Gy. The mean values of D 99% and D 80% for IDL 50% were 44.3 and 51.9 Gy.

Conclusion

The optimal IDL is 40%–50% for GTV > 0.1 cm 3 . These lower IDLs could increase D 99% and D 80% of GTV while lowering V 25Gy of normal brain tissue, which may help reduce the risk of radiation necrosis and improve local control.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Brain NeoplasmsRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiosurgeryOrgans at RiskRadiotherapy, Intensity-ModulatedHumansRetrospective StudiesParticle AcceleratorsMaleFemalePrognosisMiddle AgedAged

Stichwörter der Autor:innen

HyperArcbrain metastasesisodose linestereotactic radiotherapytreatment planning

Autor:innen und Institutionen

  • Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan.

ORCID-Profile: Tomohiro Sagawa, Toshiki Ikawa, Naoyuki Kanayama, Shoki Inui

Zitieren

APA

Sagawa, T., Ikawa, T., Ohira, S., Kanayama, N., Ueda, Y., Inui, S., Miyazaki, M., & Konishi, K. (2024). What is the optimal isodose line for stereotactic radiotherapy for single brain metastases using HyperArc?. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14408

Vancouver

Sagawa T, Ikawa T, Ohira S, Kanayama N, Ueda Y, Inui S, et al. What is the optimal isodose line for stereotactic radiotherapy for single brain metastases using HyperArc?. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14408

BibTeX
@article{Sagawa2024,
  author = {Sagawa, Tomohiro and Ikawa, Toshiki and Ohira, Shingo and Kanayama, Naoyuki and Ueda, Yoshihiro and Inui, Shoki and Miyazaki, Masayoshi and Konishi, Koji},
  title = {What is the optimal isodose line for stereotactic radiotherapy for single brain metastases using HyperArc?},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14408}
}
Kostenlos starten →