Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14343 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Leon Dunn und 4 weiteren Autor:innen. Beteiligte Einrichtung: GenesisCare Berwick, St John of God Berwick Specialist Centre, Berwick, Victoria, Australia. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Brain Neoplasms, Radiosurgery. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Single‐isocenter multi‐target intracranial stereotactic radiotherapy (SIMT) is an effective treatment for brain metastases with complex treatment plans and delivery optimization necessitating rigorous quality assurance. This work aims to assess five methods for quality assurance of SIMT treatment plans in terms of their suitability and sensitivity to delivery errors.
Methods
Sun Nuclear ArcCHECK and SRS MapCHECK, GafChromic EBT Radiochromic Film, machine log files, and Varian Portal Dosimetry were all used to measure 15 variations of a single SIMT plan. Variations of the original plan were created with Python. They comprised various degrees of systematic MLC offsets per leaf up to 2 mm, random per‐leaf variations with differing minimum and maximum magnitudes, simulated collimator, and dose miscalibrations (MU scaling). The erroneous plans were re‐imported into Eclipse and plan‐quality degradation was assessed by comparing each plan variation to the original clinical plan in terms of the percentage of clinical goals passing relative to the original plan. Each erroneous plan could be then ranked by the plan‐quality degradation percentage following recalculation in the TPS so that the effects of each variation could be correlated with γ pass rates and detector suitability. Results & conclusions It was found that 2%/1 mm is a good starting point for the ArcCHECK, Portal Dosimetry, and the SRS MapCHECK methods, respectively, and provides clinically relevant error detection sensitivity. Looser dose criteria of 5%/1 mm or 5%/1.5 mm are suitable for film dosimetry and log‐file‐based methods. The statistical methods explored can be expanded to other areas of patient‐specific QA and detector assessment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- GenesisCare Berwick, St John of God Berwick Specialist Centre, Berwick, Victoria, Australia.
- GenesisCare Frankston, Frankston Public Surgical Centre, Frankston, Victoria, Australia.
- GenesisCare Ringwood, Ringwood Private Hospital, Ringwood East, Victoria, Australia.
Zitieren
APA
Dunn, L., Tamborriello, A., Subramanian, B., Xu, X., & Ruruku, T. T. (2024). Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14343
Vancouver
Dunn L, Tamborriello A, Subramanian B, Xu X, Ruruku TT. Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14343
BibTeX
@article{Dunn2024,
author = {Dunn, Leon and Tamborriello, Alessia and Subramanian, Brindha and Xu, Xiaolei and Ruruku, Tyrone Te},
title = {Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14343}
}
EUVIMED