Systematic evaluation of spatial resolution and gamma criteria for quality assurance with detector arrays in stereotactic radiosurgery
Journal of Applied Clinical Medical Physics · 2024-02 · 25 (2)
DOI: 10.1002/acm2.14274 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 24.01.2024. Verfasst von Ann-Kathrin Stedem und 5 weiteren Autor:innen. Beteiligte Einrichtung: Asklepios Klinik St. Georg, Hermann-Holthusen-Institut für Strahlentherapie, Hamburg, Germany. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To characterize detector array spacing and gamma index for quality assurance (QA) of stereotactic radiosurgery (SRS) deliveries. Use the Nyquist theorem to determine the required detector spacing in SRS fields, and find optimal gamma indices to detect MLC errors using the SRS MapCHECK, ArcCHECK, and a portal imaging device (EPID).
Methods
The required detector spacing was determined via Fourier analysis of small radiation fields and profiles of typical SRS treatment plans. The clinical impact of MLC errors of 0.5, 1, and 2 mm was evaluated. Global gamma (low‐dose threshold 10%) was evaluated for the three detector systems using various combinations of the distance to agreement and the dose difference.
Results
While MLC errors only slightly affected mean dose to PTV and a 2 mm thick surrounding structure (PTV_2 mm), significant PTV underdose incurred with increase in maximum dose to PTV_2 mm. Gamma indices with highest sensitivity to the introduced errors at 95% tolerance level for plans on target volumes of 3.2 cm 3 (plan 3 cc) and 35.02 cm 3 (plan 35 cc) were 2%/1 mm for the SRS MapCHECK and 2%/3 mm for the ArcCHECK, with 3%/1 mm (plan 3cc) and 2%/1 mm (plan 35cc) for the EPID. Drops in passing rates for a 2 mm MLC error were (46.2%, 41.6%) for the SRS MapCHECK and (12.2%, 4.2%) for the ArcCHECK for plan 3cc and plan 35cc, respectively. For Portal Dose, values were 4.5% (plan 3cc) and 7% (plan 35cc). The Nyquist frequency of two SRS dose distributions lie between 0.26 and 0.1 mm −1 , corresponding to detector spacings of 1.9 and 5 mm. Evaluation of SRS MapCHECK data with doubled detector density indicates that increased detector density may reduce the system's sensitivity to errors, necessitating a tighter gamma index.
Conclusions
The present results give insight on the performance of detector arrays and gamma indices for the investigated detectors during SRS QA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Asklepios Klinik St. Georg, Hermann-Holthusen-Institut für Strahlentherapie, Hamburg, Germany.
- Radiotherapy Department, Beacon Hospital, Dublin, Ireland.
- Physics Research and Outreach, Sun Nuclear, A Mirion Medical Company, Norderstedt, Germany.
- Klinik für Strahlentherapie, Wilhelmshaven, Germany.
- National University of Ireland, School of Physics, University Road, Galway, Ireland.
ORCID-Profile: Ndimofor Chofor
Zitieren
APA
Stedem, A. K., Tutty, M., Chofor, N., Langhans, M., Kleefeld, C., & Schönfeld, A. A. (2024). Systematic evaluation of spatial resolution and gamma criteria for quality assurance with detector arrays in stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 25(2). https://doi.org/10.1002/acm2.14274
Vancouver
Stedem AK, Tutty M, Chofor N, Langhans M, Kleefeld C, Schönfeld AA. Systematic evaluation of spatial resolution and gamma criteria for quality assurance with detector arrays in stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2024;25(2). doi:10.1002/acm2.14274
BibTeX
@article{Stedem2024,
author = {Stedem, Ann-Kathrin and Tutty, Mark and Chofor, Ndimofor and Langhans, Marco and Kleefeld, Christoph and Schönfeld, Andreas A},
title = {Systematic evaluation of spatial resolution and gamma criteria for quality assurance with detector arrays in stereotactic radiosurgery},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {2},
doi = {10.1002/acm2.14274}
}
EUVIMED