EUVIMEDEUVIMED entdecken
Crossref

Improving Lung Stereotactic Body Radiation Therapy Dose Conformity Using a Simple Noncoplanar Volumetric Modulated Arc Therapy Technique

Tiffany Kei, Kirk Luca, Oluwatosin Kayode, Kristin A Higgins, Jeffrey D Bradley, Joseph W Shelton, Ashish B Patel, Xiaofeng Yang, Eduard Schreibmann, Jiahan Zhang, Aparna H Kesarwala, Justin Roper

Practical Radiation Oncology · 2025

DOI: 10.1016/j.prro.2025.02.007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Practical Radiation Oncology, 2025. Verfasst von Tiffany Kei und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia; Medical Dosimetry Program, College of Health and Human Scien…. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated, Lung Neoplasms, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben. Gefördert von: American Cancer Society, National Institutes of Health, National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study characterizes lung stereotactic body radiation therapy (SBRT) dose conformity for a noncoplanar volumetric modulated arc therapy (VMAT) technique. Retrospectively, 288 previously treated lung SBRT cases were replanned using a VMAT technique of 2 ipsilateral 180° arcs separated by 30° couch angles. Two objectives were used for optimization: a lower objective to achieve planning target volume (PTV) coverage and a custom normal tissue objective to steepen the dose gradient. The dose was calculated using Acuros. PTV coverage was 95%. Doses to the spinal cord, chest wall, esophagus, great vessels, heart, lungs, and trachea were evaluated. Conformity index (CI, isodose volume/PTV) values were recorded at the 10% to 100% isodose levels. CI50% results were benchmarked against the corresponding clinical plans and evaluated using the Wilcoxon signed-rank test. Linear regression was performed to characterize the relationship between dose conformity and the following PTV features: Hounsfield Units, volume, surface area, surface-to-volume ratio, and compactness. Compared with the clinical plans, the 2-objective VMAT plans demonstrated comparable or superior sparing of organs at risk with improvements in CI at the 10% to 100% isodose levels, all of which were statistically significant (P < .001). The average reductions in CI30% and CI50% were 3.5 and 0.63, respectively. Compared with the clinical plans, cases exceeding Radiation Therapy Oncology Group CI50% limits were reduced from n = 10 to 0 unacceptable and n = 78 to 20 acceptable variations. At CI30% to CI60%, regression showed that PTV Hounsfield Units and surface-to-volume ratio were significant (P < .001) predictors of dose conformity. An easily implementable VMAT technique achieved improved conformity across a broad range of lung SBRT cases and is now the standard at our institution. Further, dose conformity was characterized at different isodose levels with consideration of PTV features. Results from this study supplement historic clinical trial guidelines by providing more comprehensive and patient-specific goals for lung SBRT dose conformity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy, Intensity-ModulatedLung NeoplasmsRadiotherapy Planning, Computer-AssistedHumansRetrospective StudiesRadiotherapy DosageOrgans at Risk

Autor:innen und Institutionen

  • Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia; Medical Dosimetry Program, College of Health and Human Sciences, Southern Illinois University Carbondale, Carbondale, Illinois.
  • Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia.
  • Department of Radiation Oncology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania.
  • Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia. Electronic address: aparna.kesarwala@emory.edu.
  • Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia. Electronic address: justin.roper@emory.edu.

Förderung

  • American Cancer Society
  • National Institutes of Health
  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Kei, T., Luca, K., Kayode, O., Higgins, K. A., Bradley, J. D., Shelton, J. W., Patel, A. B., Yang, X., Schreibmann, E., Zhang, J., Kesarwala, A. H., & Roper, J. (2025). Improving Lung Stereotactic Body Radiation Therapy Dose Conformity Using a Simple Noncoplanar Volumetric Modulated Arc Therapy Technique. Practical Radiation Oncology. https://doi.org/10.1016/j.prro.2025.02.007

Vancouver

Kei T, Luca K, Kayode O, Higgins KA, Bradley JD, Shelton JW, et al. Improving Lung Stereotactic Body Radiation Therapy Dose Conformity Using a Simple Noncoplanar Volumetric Modulated Arc Therapy Technique. Practical Radiation Oncology. 2025. doi:10.1016/j.prro.2025.02.007

BibTeX
@article{Kei2025,
  author = {Kei, Tiffany and Luca, Kirk and Kayode, Oluwatosin and Higgins, Kristin A and Bradley, Jeffrey D and Shelton, Joseph W and Patel, Ashish B and Yang, Xiaofeng and Schreibmann, Eduard and Zhang, Jiahan and Kesarwala, Aparna H and Roper, Justin},
  title = {Improving Lung Stereotactic Body Radiation Therapy Dose Conformity Using a Simple Noncoplanar Volumetric Modulated Arc Therapy Technique},
  journal = {Practical Radiation Oncology},
  year = {2025},
  doi = {10.1016/j.prro.2025.02.007}
}
Kostenlos starten →