SGRT‐based stereotactic body radiotherapy for lung cancer setup accuracy and margin of the PTV

Hai-Liang Guo, Wei-Wei Wu, Yan Huan, Huai-Wen Zhang

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14195 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Hai-Liang Guo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology, the First Affiliated Hospital of Gannan Medical University, Ganzhou, China. Hauptthemen laut MEDLINE: Lung Neoplasms, Radiosurgery, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Surface‐guided radiation therapy (SGRT, AlignRT) was used to analyze motion during stereotactic body radiotherapy (SBRT) in lung cancer patients and to explore the margin of the planning target volume (PTV).

Methods

The residual errors of the AlignRT were evaluated based on grayscale cone‐beam computed tomography registration results before each treatment. AlignRT log file was used to analyze the correlation between the frequency and longest duration of errors larger than 2 mm and lasting longer than 2 s and maximum error with age and treatment duration. The displacement value at the end of treatment, the average displacement value, and the 95% probability density displacement interval were defined as intrafraction errors, and PTV1, PTV2, PTV3 were calculated by Van Herk formula or Z score analysis. Organ dosimetric differences were compared after the experience‐based margin was replaced with PTV3.

Results

The interfraction residual errors were Vrt 0 , 0.06 ± 0.18 cm; Lng 0 , ‐0.03 ± 0.19 cm; Lat 0 , 0.02 ± 0.15 cm; Pitch 0 , 0.23 ± 0.7°; Roll 0 , 0.1 ± 0.69°; Rtn 0 , ‐0.02 ± 0.79°. The frequency, longest duration and maximum error in vertical direction were correlated with treatment duration ( r = 0.404, 0.353, 0.283, p < 0.05, respectively). In the longitudinal direction, the frequency was correlated with age and treatment duration ( r = 0.376, 0.283, p < 0.05, respectively), maximum error was correlated with age ( r = 0.4, P < 0.05). Vertical, longitudinal, lateral margins of PTV1, PTV2, PTV3 were 2 mm, 4 mm, 2 mm; 2 mm, 2 mm, 2 mm, 3 mm, 5 mm, 3 mm, respectively. After replacing the original PTV, mean lung dose (MLD), 2‐cm 3 chest wall dose (CD), lung V 20 decreased by 0.2 Gy, 2.1 Gy, 0.5%, respectively ( p < 0.05).

Conclusion

AlignRT can be used for interfraction setup and monitoring intrafraction motion. It is more reasonable to use upper and lower limits of the 95% probability density interval as an intrafraction error.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lung NeoplasmsRadiosurgeryRadiotherapy, Image-GuidedHumansLungCone-Beam Computed TomographyRadiotherapy Planning, Computer-AssistedRadiotherapy Setup Errors

Stichwörter der Autor:innen

intrafraction errorlung cancerstereotactic radiotherapysurface-guided radiation therapy

Autor:innen und Institutionen

  • Department of Oncology, the First Affiliated Hospital of Gannan Medical University, Ganzhou, China.
  • Department of Radiotherapy, the Affiliated Cancer Hospital of Gannan Medical University, GanZhou Cancer Hospital, Ganzhou, China.
  • Department of Oncology, People's Hospital of Qianxinan Buyi and Miao Minority Autonomous Prefecture, Qian xinan, China.
  • Department of Radiotherapy, Jiangxi Cancer Hospital, The Second Affiliated Hospital of Nanchang Medical College, NHC Key Laboratory of Personalized Diagnosis and Treatment of Nasopharyngeal Carcinoma, Nanchang, China.
  • Department of Oncology, The Third People's Hospital of Jingdezhen, Jingdezhen, China.

Zitieren

APA

Guo, H. L., Wu, W. W., Huan, Y., & Zhang, H. W. (2024). SGRT‐based stereotactic body radiotherapy for lung cancer setup accuracy and margin of the PTV. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14195

Vancouver

Guo HL, Wu WW, Huan Y, Zhang HW. SGRT‐based stereotactic body radiotherapy for lung cancer setup accuracy and margin of the PTV. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14195

BibTeX
@article{Guo2024,
  author = {Guo, Hai-Liang and Wu, Wei-Wei and Huan, Yan and Zhang, Huai-Wen},
  title = {SGRT‐based stereotactic body radiotherapy for lung cancer setup accuracy and margin of the PTV},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14195}
}
Kostenlos starten →