A comprehensive evaluation of advanced dose calculation algorithms for brain stereotactic radiosurgery
Journal of Applied Clinical Medical Physics · 2023-11 · 24 (11)
DOI: 10.1002/acm2.14169 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.09.2023. Verfasst von Jihyung Yoon und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Rochester Medical Center, Rochester, New York, USA. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Purpose
Accurate dose calculation is important in both target and low dose normal tissue regions for brain stereotactic radiosurgery (SRS). In this study, we aim to evaluate the dosimetric accuracy of the two advanced dose calculation algorithms for brain SRS.
Methods
Retrospective clinical case study and phantom study were performed. For the clinical study, 138 SRS patient plans (443 targets) were generated using BrainLab Elements Voxel Monte Carlo (VMC). To evaluate the dose calculation accuracy, the plans were exported into Eclipse and recalculated with Acuros XB (AXB) algorithm with identical beam parameters. The calculated dose at the target center (Dref), dose to 95% target volume (D95), and the average dose to target (Dmean) were compared. Also, the distance from the skull was analyzed. For the phantom study, a cylindrical phantom and a head phantom were used, and the delivered dose was measured by an ion chamber and EBT3 film, respectively, at various locations. The measurement was compared with the calculated doses from VMC and AXB.
Results
In clinical cases, VMC dose calculations tended to be higher than AXB. It was found that the difference in Dref showed > 5% in some cases for smaller volumes < 0.3 cm 3 . Dmean and D95 differences were also higher for small targets. No obvious trend was found between the dose difference and the distance from the skull. In phantom studies, VMC dose was also higher than AXB for smaller targets, and VMC showed better agreement with the measurements than AXB for both point dose and high dose spread.
Conclusion
The two advanced calculation algorithms were extensively compared. For brain SRS, AXB sometimes calculates a noticeable lower target dose for small targets than VMC, and VMC tends to have a slightly closer agreement with measurements than AXB.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Rochester Medical Center, Rochester, New York, USA.
ORCID-Profile: Sean M Tanny
Förderung
- NCI NIH HHS
Zitieren
APA
Yoon, J., Jung, H., Tanny, S. M., Lemus, O. M. D., Milano, M. T., Hardy, S. J., Usuki, K. Y., & Zheng, D. (2023). A comprehensive evaluation of advanced dose calculation algorithms for brain stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 24(11). https://doi.org/10.1002/acm2.14169
Vancouver
Yoon J, Jung H, Tanny SM, Lemus OMD, Milano MT, Hardy SJ, et al. A comprehensive evaluation of advanced dose calculation algorithms for brain stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2023;24(11). doi:10.1002/acm2.14169
BibTeX
@article{Yoon2023,
author = {Yoon, Jihyung and Jung, Hyunuk and Tanny, Sean M and Lemus, Olga Maria Dona and Milano, Michael T and Hardy, Sara J and Usuki, Kenneth Y and Zheng, Dandan},
title = {A comprehensive evaluation of advanced dose calculation algorithms for brain stereotactic radiosurgery},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
volume = {24},
number = {11},
doi = {10.1002/acm2.14169}
}
EUVIMED