Evaluation and correlation of patient movement during online adaptive radiotherapy with CBCT and a surface imaging system

Dennis N Stanley iD, Elizabeth Covington, Joseph Harms, Joel Pogue, Carlos E Cardenas, Richard A Popple iD

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14133 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Dennis N Stanley und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama, USA. Hauptthemen laut MEDLINE: Radiotherapy, Image-Guided, Spiral Cone-Beam Computed Tomography. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

With the clinical implementation of kV‐CBCT‐based daily online‐adaptive radiotherapy, the ability to monitor, quantify, and correct patient movement during adaptive sessions is paramount. With sessions lasting between 20–45 min, the ability to detect and correct for small movements without restarting the entire session is critical to the adaptive workflow and dosimetric outcome. The purpose of this study was to quantify and evaluate the correlation of observed patient movement with machine logs and a surface imaging (SI) system during adaptive radiation therapy.

Methods

Treatment machine logs and SGRT registration data log files for 1972 individual sessions were exported and analyzed. For each session, the calculated shifts from a pre‐delivery position verification CBCT were extracted from the machine logs and compared to the SGRT registration data log files captured during motion monitoring. The SGRT calculated shifts were compared to the reported shifts of the machine logs for comparison for all patients and eight disease site categories.

Results

The average (±STD) net displacement of the SGRT shifts were 2.6 ± 3.4 mm, 2.6 ± 3.5 mm, and 3.0 ± 3.2 in the lateral, longitudinal, and vertical directions, respectively. For the treatment machine logs, the average net displacements in the lateral, longitudinal, and vertical directions were 2.7 ± 3.7 mm, 2.6 ± 3.7 mm, and 3.2 ± 3.6 mm. The average difference (Machine–SGRT) was −0.1 ± 1.8 mm, 0.2 ± 2.1 mm, and −0.5 ± 2.5 mm for the lateral, longitudinal, and vertical directions. On average, a movement of 5.8 ± 5.6 mm and 5.3 ± 4.9 mm was calculated prior to delivery for the CBCT and SGRT systems, respectively. The Pearson correlation coefficient between CBCT and SGRT shifts was r = 0.88. The mean and median difference between the treatment machine logs and SGRT log files was less than 1 mm for all sites.

Conclusion

Surface imaging should be used to monitor and quantify patient movement during adaptive radiotherapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Image-GuidedSpiral Cone-Beam Computed TomographyHumansPatient PositioningRadiotherapy Planning, Computer-AssistedMovementRadiotherapy DosageCone-Beam Computed Tomography

Stichwörter der Autor:innen

IDENTIFYSGRTonline adaptive radiotherapyoptical surface imagingpatient movement

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama, USA.
  • Department of Radiation Oncology, Michigan Medicine, Ann Arbor, Michigan, USA.

ORCID-Profile: Dennis N Stanley, Richard A Popple

Förderung

  • NCI NIH HHS

Zitieren

APA

Stanley, D. N., Covington, E., Harms, J., Pogue, J., Cardenas, C. E., & Popple, R. A. (2023). Evaluation and correlation of patient movement during online adaptive radiotherapy with CBCT and a surface imaging system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14133

Vancouver

Stanley DN, Covington E, Harms J, Pogue J, Cardenas CE, Popple RA. Evaluation and correlation of patient movement during online adaptive radiotherapy with CBCT and a surface imaging system. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14133

BibTeX
@article{Stanley2023,
  author = {Stanley, Dennis N and Covington, Elizabeth and Harms, Joseph and Pogue, Joel and Cardenas, Carlos E and Popple, Richard A},
  title = {Evaluation and correlation of patient movement during online adaptive radiotherapy with CBCT and a surface imaging system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14133}
}
Kostenlos starten →