Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy

Takahiro Kato, Sho Sasaki, Tomohiro Ikeda iD, Ryohei Kato, Masato Kato iD, Yuki Narita, Sho Oyama, Shinya Komori, Takaomi Harada, Masao Murakami

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14043 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Takahiro Kato und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan. Hauptthemen laut MEDLINE: Proton Therapy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the dosimetric effect of six degrees of freedom (6DoF) couch top with rotational corrections in proton therapy (PT).

Methods

The water equivalent thickness (WET) was measured using a proton beam with a 6DoF couch top and patient immobilization base plate (PIBP) placed in front of a motorized water phantom. The accuracy verification was performed with the beam axis set perpendicular to the 6DoF couch top and tilted in 10° steps from 10° to 30°. Up to 3° rotational correction may be added during the actual treatment to correct the rotational setup error on our system. The measured and calculated values using the treatment planning system were compared. Additionally, the effect of the 3° difference was evaluated using actual measurements concerning each angle on the proton beam range.

Results

The WET of the 6DoF couch top and PIBP were 8.5 ± 0.1 mm and 6.8 ± 0.1 mm, respectively. The calculation and the actual measurement at each angle agreed within 0.2 mm at the maximum. A maximum difference of approximately 0.6 mm was confirmed when tilted at 3° following 30° with the 6DoF couch top plus PIBP.

Conclusions

The dosimetric effect of the 6DoF couch top with rotational corrections in PT differs depending on the incidence angle on the couch top, and it increased with the increased oblique angle of incidence. However, the effect on the range was as small as 0.6 mm at the maximum. The amount of rotational correction, the angle of incidence of the beam, and the effect of rotational corrections on the proton beam range may differ depending on the structure of the couch top. Therefore, sufficient prior confirmation, and subsequent periodical quality assurance management are important.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyHumansPatient PositioningProtonsRadiometryRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

proton beam rangeproton therapysix degrees of freedom couch

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan.
  • Department of Radiation Physics and Technology, Southern Tohoku Proton Therapy Center, Fukushima, Japan.
  • Department of Radiation Oncology, Southern Tohoku Proton Therapy Center, Fukushima, Japan.

ORCID-Profile: Tomohiro Ikeda, Masato Kato

Zitieren

APA

Kato, T., Sasaki, S., Ikeda, T., Kato, R., Kato, M., Narita, Y., Oyama, S., Komori, S., Harada, T., & Murakami, M. (2023). Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14043

Vancouver

Kato T, Sasaki S, Ikeda T, Kato R, Kato M, Narita Y, et al. Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14043

BibTeX
@article{Kato2023,
  author = {Kato, Takahiro and Sasaki, Sho and Ikeda, Tomohiro and Kato, Ryohei and Kato, Masato and Narita, Yuki and Oyama, Sho and Komori, Shinya and Harada, Takaomi and Murakami, Masao},
  title = {Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14043}
}
Kostenlos starten →