Evaluation of treatment planning system accuracy in estimating the stopping‐power ratio of immobilization devices for proton therapy

Kai Jiang, Michael MacFarlane, Sina Mossahebi, Mark J Zakhary

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.13831 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Kai Jiang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: Proton Therapy. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess treatment planning system (TPS) accuracy in estimating the stopping‐power ratio (SPR) of immobilization devices commonly used in proton therapy and to evaluate the dosimetric effect of SPR estimation error for a set of clinical treatment plans.

Methods

Computed tomography scans of selected clinical immobilization devices were acquired. Then, the water‐equivalent thickness (WET) and SPR values of these devices based on the scans were estimated in a commercial TPS. The reference SPR of each device was measured using a multilayer ion chamber (MLIC), and the differences between measured and TPS‐estimated SPRs were calculated. These findings were utilized to calculate corrected dose distributions of 15 clinical proton plans for three treatment sites: extremity, abdomen, and head‐and‐neck. The original and corrected dose distributions were compared using a set of target and organs‐at‐risk (OARs) dose–volume histogram (DVH) parameters.

Results

On average, the TPS‐estimated SPR was 19.5% lower (range, −35.1% to 0.2%) than the MLIC‐measured SPR. Due to the relatively low density of most immobilization devices used, the WET error was typically

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyHumansProtonsRadiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedRadiometry

Stichwörter der Autor:innen

immobilizationproton radiotherapystopping power ratiotreatment planning systemwater equivalent thickness

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland, USA.

Zitieren

APA

Jiang, K., MacFarlane, M., Mossahebi, S., & Zakhary, M. J. (2023). Evaluation of treatment planning system accuracy in estimating the stopping‐power ratio of immobilization devices for proton therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13831

Vancouver

Jiang K, MacFarlane M, Mossahebi S, Zakhary MJ. Evaluation of treatment planning system accuracy in estimating the stopping‐power ratio of immobilization devices for proton therapy. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13831

BibTeX
@article{Jiang2023,
  author = {Jiang, Kai and MacFarlane, Michael and Mossahebi, Sina and Zakhary, Mark J},
  title = {Evaluation of treatment planning system accuracy in estimating the stopping‐power ratio of immobilization devices for proton therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.13831}
}
Kostenlos starten →