Technical note: Consistency of IAEA's TRS‐483 and AAPM's extended TG‐51 protocols for clinical reference dosimetry of the CyberKnife M6 machine

Jasmine Duchaine, Daniel Markel, Jean-Luc Ley, Dominique Béliveau-Nadeau, Karim Zerouali, Robert Doucet, Hugo Bouchard

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.13976 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Jasmine Duchaine und 6 weiteren Autor:innen. Beteiligte Einrichtung: Département de physique, Université de Montréal, Campus MIL, Montréal, Québec, Canada. Hauptthemen laut MEDLINE: Photons, Phenylpropionates. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

While IAEA's TRS‐483 code of practice is adapted for the calibration of CyberKnife machines, AAPM's TG‐51 is still the protocol recommended by the manufacturer for their calibration. The differences between both protocols could lead to differences in absorbed dose to water during the calibration process.

Purpose

The aims of this work are to evaluate the difference resulting from the application of TG‐51 (including the manufacturer's adaptations) and TRS‐483 in terms of absorbed dose to water for a CyberKnife M6, and to evaluate the consistency of TRS‐483.

Methods

Measurements are performed on a CyberKnife M6 unit under machine‐specific reference conditions using a calibrated Exradin A12 ionization chamber. Monte Carlo (MC) simulations are performed to estimate and using a fully modeled detector and an optimized CyberKnife M6 beam model. The latter is also estimated experimentally. Differences between the adapted TG‐51 and TRS‐483 protocols are identified and their impact is quantified.

Results

When using an in‐house experimentally‐evaluated volume averaging correction factor, a difference of 0.11% in terms of absorbed dose to water per monitor unit is observed when applying both protocols. This disparity is solely associated to the difference in beam quality correction factor. If a generic volume averaging correction factor is used during the application of TRS‐483, the difference in calibration increases to 0.14%. In both cases, the disparity is not statistically significant according to TRS‐483's reported uncertainties on their beam quality correction factor (i.e., 1%). MC results lead to and . Results illustrate that the generic beam quality correction factor provided in the TRS‐483 might be overestimated by 0.36% compared to our specific model and that this overestimation could be due to the volume averaging component.

Conclusions

For clinical reference dosimetry of the CyberKnife M6, the application of TRS‐483 is found to be consistent with TG‐51.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotonsPhenylpropionatesHumansRadiometryMonte Carlo MethodWaterCalibration

Stichwörter der Autor:innen

CyberKnifeTG-51TRS-483reference dosimetry

Substanzen

TEI 5103PhenylpropionatesWater

Autor:innen und Institutionen

  • Département de physique, Université de Montréal, Campus MIL, Montréal, Québec, Canada.
  • Centre de recherche du Centre hospitalier de l'Université de Montréal, Montréal, Québec, Canada.
  • Département de radio-oncologie, Centre hospitalier de l'Université de Montréal, Montréal, Québec, Canada.

Förderung

  • Natural Sciences and Engineering Research Council of Canada

Zitieren

APA

Duchaine, J., Markel, D., Ley, J. L., Béliveau-Nadeau, D., Zerouali, K., Doucet, R., & Bouchard, H. (2023). Technical note: Consistency of IAEA's TRS‐483 and AAPM's extended TG‐51 protocols for clinical reference dosimetry of the CyberKnife M6 machine. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13976

Vancouver

Duchaine J, Markel D, Ley JL, Béliveau-Nadeau D, Zerouali K, Doucet R, et al. Technical note: Consistency of IAEA's TRS‐483 and AAPM's extended TG‐51 protocols for clinical reference dosimetry of the CyberKnife M6 machine. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13976

BibTeX
@article{Duchaine2023,
  author = {Duchaine, Jasmine and Markel, Daniel and Ley, Jean-Luc and Béliveau-Nadeau, Dominique and Zerouali, Karim and Doucet, Robert and Bouchard, Hugo},
  title = {Technical note: Consistency of IAEA's TRS‐483 and AAPM's extended TG‐51 protocols for clinical reference dosimetry of the CyberKnife M6 machine},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.13976}
}
Kostenlos starten →