A Comparative Phantom Study: Effectiveness Of A NIM PET/CT Phantom To Evaluate PET Image Quality In Detection Of Micro-Lesions Comparing With NEMA IEC Body Phantom And Computed Tomography Performance Parameters
IOSR Journal of Dental and Medical Sciences · 2026-05 · 25 (5) · 01-08
DOI: 10.9790/0853-2505010108 ↗Kurz gefasst
Fachartikel in der Zeitschrift IOSR Journal of Dental and Medical Sciences, 2026-05. Verfasst von Rasool Kattubadi Bhujeet Kumar und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Background
This comparative study evaluated the PET image quality with commonly used NEMA IEC Body and observed a number of defects, hindering its application for detecting micro-lesions and measuring the performance parameters of computed tomography (CT). This study aimed to compare, review, and propose a PET/CT phantom designed by National Institute of Metrology (NIM), China, and that phantom tested in our organisation, which is capable of simultaneously testing the performance of PET and CT systems, and to evaluate the quality of imaging. Initial NIM study was published in BMC medical imaging in 2021 and further we did this comparative study to review the details more thoroughly in 2025.
Methods
The phantom was developed in the current study, the NIM PET/CT phantom, is composed of a PET imaging module and a CT imaging module, and these modules are connected together through bolts, which can simul- tenuously measure the imaging performance of PET and CT systems and similar to NEMA IEC Body phantom. Hot spheres were filled with 4:1 sphere-to- background activity concentration using 18F-fluorodeoxyglucose (18F-FDG), and cold spheres were filled with non- radioactive water. The results of imaging obtained from the NIM PET/CT phantom and the NEMA IEC Body phantom were compared to assess their diagnostic efficacy. In order to evaluate the generalization ability of the NIM PET/CT phantom, three different PET/CT systems were used to scan on the same scanning protocol. To evaluate the effects of image reconstruction algorithms on image quality assessment, VPFX (Vue point FX), VPHD (Vue point HD), ordered subset expectation maximization (OSEM), OSEM-point-spread function (PSF), OSEM-TOF, and OSEM-PSF-TOF algorithms were employed using Discovery MI Gen2, 3 ring scanner (GE).
Results
The imaging quality of the NIM PET/CT phantom and the NEMA IEC Body phantom was relatively consistent. The NIM PET/CT phantom could detect 7 mm spheres without influencing the imaging quality. It was found that PSF reconstruction exhibited to reduce the speed of convergence, the contrast and background variability of spheres (13 –28 mm) were significantly improved after two iterations. In addition to improve the image contrast and background variability, TOF could markedly improve the overall image quality and instrument detection limit. TOF-PSF could noticeably reduce noise level, enhance imaging details, and improve quality of imaging.
Conclusions
The results showed that in comparison with the NEMA IEC Body phantom, the NIM PET/CT phantom outperformed in evaluating the PET image quality of micro-lesions and the performance parameters of CT
Zitieren
APA
Kumar, R. K. B., & Mohamedahmed, E. (2026). A Comparative Phantom Study: Effectiveness Of A NIM PET/CT Phantom To Evaluate PET Image Quality In Detection Of Micro-Lesions Comparing With NEMA IEC Body Phantom And Computed Tomography Performance Parameters. IOSR Journal of Dental and Medical Sciences, 25(5), 01-08. https://doi.org/10.9790/0853-2505010108
Vancouver
Kumar RKB, Mohamedahmed E. A Comparative Phantom Study: Effectiveness Of A NIM PET/CT Phantom To Evaluate PET Image Quality In Detection Of Micro-Lesions Comparing With NEMA IEC Body Phantom And Computed Tomography Performance Parameters. IOSR Journal of Dental and Medical Sciences. 2026;25(5):01-08. doi:10.9790/0853-2505010108
BibTeX
@article{Kumar2026,
author = {Kumar, Rasool Kattubadi Bhujeet and Mohamedahmed, Elmubarak},
title = {A Comparative Phantom Study: Effectiveness Of A NIM PET/CT Phantom To Evaluate PET Image Quality In Detection Of Micro-Lesions Comparing With NEMA IEC Body Phantom And Computed Tomography Performance Parameters},
journal = {IOSR Journal of Dental and Medical Sciences},
year = {2026},
volume = {25},
number = {5},
pages = {01--08},
doi = {10.9790/0853-2505010108}
}
EUVIMED