Electrical Impedance Changes InPediatric Cochlear Implant Recipients Among Three Different Cochlear Implant Manufacturers

Sunil Kumar, Prof. Rajesh Kumar, Prof. Vishwambhar Singh, Dr. S.K. Aggarwal, Dr. S.K. Shukla, Dr. Sanjay Bharati, Dr. Siva S Dr. Siva S

IOSR Journal of Dental and Medical Sciences · 2026-04 · 25 (4) · 49-56

DOI: 10.9790/0853-2504034956 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IOSR Journal of Dental and Medical Sciences, 2026-04. Verfasst von Sunil Kumar und 6 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Electrical impedance is a key objective parameter reflecting the electrode–tissue interface in cochlear implant (CI) recipients. Its temporal variation provides important insights into device performance and intracochlear biological responses, particularly in paediatric populations.

Objective

To evaluate and compare electrical impedance changes over time among paediatric cochlear implant recipients using three different CI manufacturers.

Methods

This prospective observational study included 30 paediatric patients (aged 2–8 years) with bilateral profound sensory-neural hearing loss that underwent cochlear implantation at a tertiary care centre. Patients were equally distributed into three groups (n=10 each) based on implant manufacturer (Company A, B, and C). Impedance measurements were recorded at four time points: intraoperative, switch-on, 3 months, and 6 months post-implantation. Statistical analysis was performed using paired t-tests and ANOVA.

Results

Intraoperative and early postoperative impedance values showed no statistically significant intergroup differences (p>0.05). However, a trend toward divergence was observed at 3 months, becoming statistically significant at 6 months (p=0.006), with Company A demonstrating higher impedance values compared to Company B. Within-group analysis revealed stable impedance in Company A, significant early increases followed by stabilization in Company B, and a marked immediate postoperative rise with subsequent plateau in Company C.

Conclusion

While early impedance values are comparable across implant systems, significant differences emerge over time, reflecting device-specific electrode characteristics and biological interactions. These findings highlight the importance of individualized device monitoring and programming in paediatric cochlear implant recipients.

Zitieren

APA

Kumar, S., Kumar, P. R., Singh, P. V., Aggarwal, D. S., Shukla, D. S., Bharati, D. S., & S, D. S. S. D. S. (2026). Electrical Impedance Changes InPediatric Cochlear Implant Recipients Among Three Different Cochlear Implant Manufacturers. IOSR Journal of Dental and Medical Sciences, 25(4), 49-56. https://doi.org/10.9790/0853-2504034956

Vancouver

Kumar S, Kumar PR, Singh PV, Aggarwal DS, Shukla DS, Bharati DS, et al. Electrical Impedance Changes InPediatric Cochlear Implant Recipients Among Three Different Cochlear Implant Manufacturers. IOSR Journal of Dental and Medical Sciences. 2026;25(4):49-56. doi:10.9790/0853-2504034956

BibTeX
@article{Kumar2026,
  author = {Kumar, Sunil and Kumar, Prof. Rajesh and Singh, Prof. Vishwambhar and Aggarwal, Dr. S.K. and Shukla, Dr. S.K. and Bharati, Dr. Sanjay and S, Dr. Siva S Dr. Siva},
  title = {Electrical Impedance Changes InPediatric Cochlear Implant Recipients Among Three Different Cochlear Implant Manufacturers},
  journal = {IOSR Journal of Dental and Medical Sciences},
  year = {2026},
  volume = {25},
  number = {4},
  pages = {49--56},
  doi = {10.9790/0853-2504034956}
}
Kostenlos starten →