EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Intraoperative and Long-Term Impedance Changes and Electrode Malfunctions in Cochlear Implant Recipients

Muhammed Dagkiran iD, Ilda Tanrisever Pehlivan iD

Journal of Clinical Medicine · 10.05.2026 · 15 (10) · 3662

DOI: 10.3390/jcm15103662 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 10.05.2026. Verfasst von Muhammed Dagkiran und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, Department of Otolaryngology and Head and Neck Surgery, Campus of Balcali, Cukurova University, 01330 Saricam, Adana, Turkey. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to investigate intraoperative and long-term impedance changes in cochlear implant recipients and to evaluate the prevalence and course of electrode circuit malfunctions.

Methods

A retrospective longitudinal observational study was conducted on 358 ears that underwent cochlear implantation between October 2023 and June 2024. Intraoperative and postoperative impedance telemetry measurements were obtained at 1, 2, 6, and 12 months. Electrode arrays were analyzed according to their positions (basal, medial, apical), and electrode malfunctions were classified as open circuit (OC) or short circuit (SC).

Results

Impedance values significantly increased from intraoperative to first activation measurements, reached their peak at the 1st postoperative month, and gradually decreased, stabilizing after the 6th month (p < 0.05). Electrode malfunctions were detected in 26 of 358 devices (7.2%), with 22 OCs (84.7%) and 4 SCs (15.3%). While 4 of the 22 OC cases improved within the first month after surgery, no improvement was observed in SC cases during the 12-month follow-up period.

Conclusions

Impedance telemetry is useful for assessing electrode integrity intraoperatively and for monitoring changes over time during follow-up. These findings highlight the dynamic nature of impedance changes, particularly during the first 6 months, and underscore the value of regular follow-up measurements and individualized programming.

Schlagwörter

Stichwörter der Autor:innen

cochlear implantelectrode malfunctionimpedance telemetryopen circuitshort circuit

Autor:innen und Institutionen

  • Faculty of Medicine, Department of Otolaryngology and Head and Neck Surgery, Campus of Balcali, Cukurova University, 01330 Saricam, Adana, Turkey.

ORCID-Profile: Muhammed Dagkiran, Ilda Tanrisever Pehlivan

Zitieren

APA

Dagkiran, M., & Tanrisever Pehlivan, I. (2026). Intraoperative and Long-Term Impedance Changes and Electrode Malfunctions in Cochlear Implant Recipients. Journal of Clinical Medicine, 15(10), 3662. https://doi.org/10.3390/jcm15103662

Vancouver

Dagkiran M, Tanrisever Pehlivan I. Intraoperative and Long-Term Impedance Changes and Electrode Malfunctions in Cochlear Implant Recipients. Journal of Clinical Medicine. 2026;15(10):3662. doi:10.3390/jcm15103662

BibTeX
@article{Dagkiran2026,
  author = {Dagkiran, Muhammed and Tanrisever Pehlivan, Ilda},
  title = {Intraoperative and Long-Term Impedance Changes and Electrode Malfunctions in Cochlear Implant Recipients},
  journal = {Journal of Clinical Medicine},
  year = {2026},
  volume = {15},
  number = {10},
  pages = {3662},
  doi = {10.3390/jcm15103662}
}
Kostenlos starten →