Evaluation of Mean Glandular Dose from Digital Mammography Exams at Qatar and Compared with International Guidelines Levels

Huda Al-Naemi, Osman Taha, Abeer Al-attar, Mahmoud Tarabieh, Ibrahim Abdallah, Nabil Iqeilan, Antar Aly

British Journal of Medicine and Medical Research · 10.01.2016 · 14 (2) · 1-9

DOI: 10.9734/bjmmr/2016/23129 ↗

Kurz gefasst

Leitlinie in der Zeitschrift British Journal of Medicine and Medical Research, online erschienen am 10.01.2016. Verfasst von Huda Al-Naemi und 6 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The primary objective of this study was to measure the mean glandular dose (MGD) from craniocaudal (CC) and mediolateral oblique (MLO) views from mammography patients in Hamad Medical Corporation (HMC) in the state of Qatar and to compare them with the international guidelines levels as well as to establish Dose Reference Level (DRL) for the country by applying the quality control (QC) protocol for the Digital mammography units and to reduce the patient dose and improve the image quality. All patients data was taken from two Selenia digital mammography units for 18 months period. Quality control was implemented for the two mammography machines and corrective actions have been done for the image quality evaluation with rejected film analysis. The total number of collected patient data was 4085 mammography exams which considered as around 93% of the overall mammography procedures done in Qatar during that period. Based on the IAEA selection criteria of breast thickness between 2-7 cm and kV machine value

Zitieren

APA

Al-Naemi, H., Taha, O., Al-attar, A., Tarabieh, M., Abdallah, I., Iqeilan, N., & Aly, A. (2016). Evaluation of Mean Glandular Dose from Digital Mammography Exams at Qatar and Compared with International Guidelines Levels. British Journal of Medicine and Medical Research, 14(2), 1-9. https://doi.org/10.9734/bjmmr/2016/23129

Vancouver

Al-Naemi H, Taha O, Al-attar A, Tarabieh M, Abdallah I, Iqeilan N, et al. Evaluation of Mean Glandular Dose from Digital Mammography Exams at Qatar and Compared with International Guidelines Levels. British Journal of Medicine and Medical Research. 2016;14(2):1-9. doi:10.9734/bjmmr/2016/23129

BibTeX
@article{AlNaemi2016,
  author = {Al-Naemi, Huda and Taha, Osman and Al-attar, Abeer and Tarabieh, Mahmoud and Abdallah, Ibrahim and Iqeilan, Nabil and Aly, Antar},
  title = {Evaluation of Mean Glandular Dose from Digital Mammography Exams at Qatar and Compared with International Guidelines Levels},
  journal = {British Journal of Medicine and Medical Research},
  year = {2016},
  volume = {14},
  number = {2},
  pages = {1--9},
  doi = {10.9734/bjmmr/2016/23129}
}
Kostenlos starten →