Breast Glandularity Distribution and Refining the Mean Glandular Dose Estimates in Digital Mammography
Journal of Medical Physics · 2025-10 · 50 (4) · 795-801
DOI: 10.4103/jmp.jmp_142_25 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Physics, online erschienen am 31.12.2025. Verfasst von Mirjeta Mediji Arifi und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Natural Sciences and Mathematics, Physics Department, Cyril and Methodius University of Skopje, Skopje, North Macedonia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
In mammography, radiation dose is typically expressed as the mean glandular dose (MGD), which represents the dose delivered to the glandular tissue of the breast. Materials and Methods: This study compares MGD estimates obtained using three different methodologies: (I) MGD-Dance-Laboratory for Individualized Breast Radiodensity Assessment (LIBRA) – Calculated manually for each patient using Dance’s formula, incorporating mammographic breast density values derived from the LIBRA application, thereby replacing Dance’s standard glandularity assumption with image-specific values; (II) MGD-Dance – Calculated using Dance’s formula with the conventional assumption of 50% glandularity; (III) MGD-Displayed – Extracted directly from the Digital Imaging and Communication in Medicine header of each mammogram.
Results
A total of 688 anonymized mammograms from 172 women undergoing routine screening were analyzed, with complete technical and patient-related data. The mean MGD values obtained by the three methods were: MGD-Dance-LIBRA: 2.97 mGy; MGD-Dance: 2.78 mGy; and MGD-Displayed: 2.81 mGy. The average glandularity across the dataset was estimated at 14%. A strong correlation was observed between MGD-Dance and MGD-Dance-LIBRA values ( R ² =0.9865). The refined dose estimation using image-specific glandularity from LIBRA consistently produced slightly higher values compared to the standard Dance method, highlighting the impact of the commonly assumed 50% glandularity, which overestimates the true average density.
Conclusions
Incorporating individualized breast density estimates from the LIBRA application into Dance’s formula provides a more refined and accurate method for calculating MGD in digital mammography.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Natural Sciences and Mathematics, Physics Department, Cyril and Methodius University of Skopje, Skopje, North Macedonia.
- Faculty of Natural Sciences and Mathematics, Physics Department, University of Tetovo, Tetovo, North Macedonia.
Zitieren
APA
Arifi, M. M., & Ristova, M. (2025). Breast Glandularity Distribution and Refining the Mean Glandular Dose Estimates in Digital Mammography. Journal of Medical Physics, 50(4), 795-801. https://doi.org/10.4103/jmp.jmp_142_25
Vancouver
Arifi MM, Ristova M. Breast Glandularity Distribution and Refining the Mean Glandular Dose Estimates in Digital Mammography. Journal of Medical Physics. 2025;50(4):795-801. doi:10.4103/jmp.jmp_142_25
BibTeX
@article{Arifi2025,
author = {Arifi, Mirjeta Mediji and Ristova, Mimoza},
title = {Breast Glandularity Distribution and Refining the Mean Glandular Dose Estimates in Digital Mammography},
journal = {Journal of Medical Physics},
year = {2025},
volume = {50},
number = {4},
pages = {795--801},
doi = {10.4103/jmp.jmp_142_25}
}