Factors influencing compliance with endoscopy final rinsing water standards: a study in a tertiary hospital setting
PeerJ · 2025
DOI: 10.7717/peerj.20134 ↗At a glance
Journal article in PeerJ, 2025. Written by Yuhua Yuan and 4 co-authors. Institution: Department of Infection Prevention and Control, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, Hangzhou, China. Major MEDLINE topics: Water Microbiology, Guideline Adherence, Infection Control, Disinfection. Cited 4 times so far (Crossref); 21 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Maintaining a high compliance rate for final rinsing water is essential for patient safety and infection control in healthcare facilities. This study aims to investigate the causes of fluctuations in compliance rates at a tertiary hospital and evaluate the effectiveness of preventive measures.
Methods
Monitoring data from October 2022 to December 2023 were analyzed to assess microbial contamination in final rinse water. Environmental assessments and literature reviews were conducted to identify potential contributing factors. Strategic interventions were implemented, and their impact on water quality and infection control was systematically evaluated.
Findings
Initial assessments revealed fluctuating compliance rates, with faucet aerators identified as a potential source of contamination. Water samples collected before the aerator showed 100% compliance (6/6 samples), while colonies of Cupriavidus pauculus and Stenotrophomonas maltophilia were detected on aerator surfaces and in final rinse water. After removing the aerators and enhancing disinfection protocols, compliance improved significantly, with subsequent samples meeting infection control standards (≤10 CFU/100 mL).
Conclusion
This study suggests a potential link between fluctuations in endoscopy final rinsing water quality and the presence of faucet aerators. These findings support further research and the development of guidelines for the appropriate use of aerators in healthcare settings.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Infection Prevention and Control, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, Hangzhou, China.
Cite
APA
Yuan, Y., Ye, L., Lu, T., Feng, B., & Zhao, J. (2025). Factors influencing compliance with endoscopy final rinsing water standards: a study in a tertiary hospital setting. PeerJ. https://doi.org/10.7717/peerj.20134
Vancouver
Yuan Y, Ye L, Lu T, Feng B, Zhao J. Factors influencing compliance with endoscopy final rinsing water standards: a study in a tertiary hospital setting. PeerJ. 2025. doi:10.7717/peerj.20134
BibTeX
@article{Yuan2025,
author = {Yuan, Yuhua and Ye, Lihong and Lu, Tianyi and Feng, Baihuan and Zhao, Jin},
title = {Factors influencing compliance with endoscopy final rinsing water standards: a study in a tertiary hospital setting},
journal = {PeerJ},
year = {2025},
doi = {10.7717/peerj.20134}
}
EUVIMED