Evaluating the Effect of Storage Temperature, Duration, and Multiple Freeze–Thaw Cycles on the RT-PCR Detection of SARS-CoV-2 RNA in Patient Samples

Taruna Singh, Prakash Tiwari, Himanshu Singh Chandel, Amaresh Nigudgi, Priyanka Singh

VINDHYA JOURNAL OF MEDICAL RESEARCH · 30.05.2026 · 2 (1)

DOI: 10.65365/vjmr.v2.i1.18 ↗

Kurz gefasst

Fachartikel in der Zeitschrift VINDHYA JOURNAL OF MEDICAL RESEARCH, online erschienen am 30.05.2026. Verfasst von Taruna Singh und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Real-time reverse transcription polymerase chain reaction (RT-PCR) is currently the standard method for detecting SARS-CoV-2 infection. Nevertheless, preanalytical issues such as storage temperature, exposure time and repeated freeze–thawing cycles can affect RNA integrity and reduce diagnostic sensitivity.

Objective

The purpose of this study was to investigate the influence of various storage temperatures, storage times, multiple freeze–thaw cycles on the stability and RT-PCR detectability of SARS-CoV-2 in clinical specimens.

Methods

A pool of nasopharyngeal and oropharyngeal swab samples obtained from SARS-CoV-2–positive patients in VTM was divided into aliquots, each submitted for storage on day 1 at different temperatures (room temperature), 4°C, −20°Cand −80°C) over time (1, 2, 3, 5 and12 days). Aliquots were frozen and thawed for various times. The viral RNA was extracted with an automatic extractor and amplified by RT-PCR to the RdRp gene. Alterations in cycle threshold (Ct) were measured.

Results

Excellent RNA stability was shown by samples kept at −80°C, with little variation in Ct values at all time periods. For up to 72 hours, storage at 4°C was acceptable; however, longer storage led to a progressive increase in Ct. Significant RNA degradation occurred after 48–72 hours of room temperature storage, especially in samples with low virus loads. Ct values increased gradually with repeated freeze-thaw cycles, and after three or more cycles, there was a noticeable loss of detectability.

Conclusion

Preanalytical storage conditions have a considerable impact on SARS-CoV-2 RT-PCR results. Maintaining RNA integrity and ensuring accurate molecular diagnosis need optimal storage at −80°C and minimizing freeze-thaw cycles.

Zitieren

APA

Singh, T., Tiwari, P., Chandel, H. S., Nigudgi, A., & Singh, P. (2026). Evaluating the Effect of Storage Temperature, Duration, and Multiple Freeze–Thaw Cycles on the RT-PCR Detection of SARS-CoV-2 RNA in Patient Samples. VINDHYA JOURNAL OF MEDICAL RESEARCH, 2(1). https://doi.org/10.65365/vjmr.v2.i1.18

Vancouver

Singh T, Tiwari P, Chandel HS, Nigudgi A, Singh P. Evaluating the Effect of Storage Temperature, Duration, and Multiple Freeze–Thaw Cycles on the RT-PCR Detection of SARS-CoV-2 RNA in Patient Samples. VINDHYA JOURNAL OF MEDICAL RESEARCH. 2026;2(1). doi:10.65365/vjmr.v2.i1.18

BibTeX
@article{Singh2026,
  author = {Singh, Taruna and Tiwari, Prakash and Chandel, Himanshu Singh and Nigudgi, Amaresh and Singh, Priyanka},
  title = {Evaluating the Effect of Storage Temperature, Duration, and Multiple Freeze–Thaw Cycles on the RT-PCR Detection of SARS-CoV-2 RNA in Patient Samples},
  journal = {VINDHYA JOURNAL OF MEDICAL RESEARCH},
  year = {2026},
  volume = {2},
  number = {1},
  doi = {10.65365/vjmr.v2.i1.18}
}
Kostenlos starten →