Accuracy of Point-of-Care Blood Glucose Meters Compared With Central Laboratory Reference Methods in Critically Ill AdultsA Systematic Review
International Journal of Health & Business Analytics · 2026
DOI: 10.65166/88wy1c14 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift International Journal of Health & Business Analytics, 2026. Verfasst von Cyrene Elijah S. Amorado und 3 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).
Abstract (Originalsprache)
Background
Point-of-care (POC) blood glucose meters shorten turnaround time for bedside glycemic management, but critical illness introduces physiologic and analytical conditions that can degrade agreement with central laboratory measurements.
Objective
To synthesize evidence on the analytical and clinical accuracy of POC blood glucose meters versus central laboratory reference methods in critically ill adults and identify conditions associated with disagreement.
Methods
Searches of PubMed, Google Scholar, and SciSpace identified 1,003 records published from 2010 through September 13, 2026. After removal of 280 duplicates, 723 unique records remained. An initial screening limit left 223 deduplicated records without title/abstract assessment; a subsequent completeness audit applied the same eligibility criteria to these records and re-evaluated full-text decisions. Thirty studies were included. Risk of bias was assessed with an adapted QUADAS-2 framework.
Results
POC performance was strongly device-, specimen-, and context-dependent. Several modern systems showed small average bias and high standards compliance in selected ICU cohorts, whereas capillary measurements deteriorated substantially in shock, hypotension, catecholamine/vasopressor exposure, anemia or hematocrit extremes, acidemia, hypoglycemia, and selected drug/interference states. Reported ISO compliance ranged from very high in stable or validated device-specific conditions to markedly poor in shock subgroups; these percentages were not directly pooled because studies used different devices, specimens, standards, thresholds, and outcome metrics.
Conclusion
POC glucose meters cannot be treated as analytically interchangeable with central laboratory testing across all critically ill adults. Risk-stratified confirmation with laboratory testing is warranted when POC results are clinically discordant or obtained under physiologic conditions known to impair meter performance.
Zitieren
APA
Amorado, C. E. S., Quinto, L., Sayas, M. K. F., & Atento, R. G. (2026). Accuracy of Point-of-Care Blood Glucose Meters Compared With Central Laboratory Reference Methods in Critically Ill AdultsA Systematic Review. International Journal of Health & Business Analytics. https://doi.org/10.65166/88wy1c14
Vancouver
Amorado CES, Quinto L, Sayas MKF, Atento RG. Accuracy of Point-of-Care Blood Glucose Meters Compared With Central Laboratory Reference Methods in Critically Ill AdultsA Systematic Review. International Journal of Health & Business Analytics. 2026. doi:10.65166/88wy1c14
BibTeX
@article{Amorado2026,
author = {Amorado, Cyrene Elijah S. and Quinto, Leah and Sayas, Mark Kerwin F. and Atento, Ramon George},
title = {Accuracy of Point-of-Care Blood Glucose Meters Compared With Central Laboratory Reference Methods in Critically Ill AdultsA Systematic Review},
journal = {International Journal of Health & Business Analytics},
year = {2026},
doi = {10.65166/88wy1c14}
}
EUVIMED