Accuracy of Noninvasive Blood Pressure Monitoring in Critically Ill Adults
Journal of Intensive Care Medicine · 2024-07 · 39 (7) · 665-671
DOI: 10.1177/08850666231225173 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Journal of Intensive Care Medicine, online erschienen am 12.01.2024. Verfasst von Erin N Haber und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pulmonary, Critical Care and Sleep Medicine, MedStar Georgetown University Hospital, Washington DC, USA. Hauptthemen laut MEDLINE: Critical Illness, Blood Pressure Determination, Intensive Care Units. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
Blood pressure (BP) is routinely invasively monitored by an arterial catheter in the intensive care unit (ICU). However, the available data comparing the accuracy of noninvasive methods to arterial catheters for measuring BP in the ICU are limited by small numbers and diverse methodologies.
Purpose
To determine agreement between invasive arterial blood pressure monitoring (IABP) and noninvasive blood pressure (NIBP) in critically ill patients.
Methods
This was a single center, observational study of critical ill adults in a tertiary care facility evaluating agreement (≤10% difference) between simultaneously measured IABP and NIBP. We measured clinical features at time of BP measurement inclusive of patient demographics, laboratory data, severity of illness, specific interventions (mechanical ventilation and dialysis), and vasopressor dose to identify particular clinical scenarios in which measurement agreement is more or less likely.
Results
Of the 1852 critically ill adults with simultaneous IABP and NIBP readings, there was a median difference of 6 mm Hg in mean arterial pressure (MAP), interquartile range (1-12), P < .01. A logistic regression analysis identified 5 independent predictors of measurement discrepancy: increasing doses of norepinephrine (adjusted odds ratio [aOR] 1.10 [95% confidence interval, CI 1.08-1.12] P = .03 for every change in 5 µg/min), lower MAP value (aOR 0.98 [0.98-0.99] P < .01 for every change in 1 mm Hg), higher body mass index (aOR 1.04 [1.01-1.09] P = .01 for an increase in 1), increased patient age (aOR 1.31 [1.30-1.37] P < .01 for every 10 years), and radial arterial line location (aOR 1.74 [1.16-2.47] P = .04).
Conclusions
There was broad agreement between IABP and NIBP in critically ill patients over a range of BPs and severity of illness. Several variables are associated with measurement discrepancy; however, their predictive capacity is modest. This may guide future study into which patients may specifically benefit from an arterial catheter.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Pulmonary, Critical Care and Sleep Medicine, MedStar Georgetown University Hospital, Washington DC, USA.
- Georgetown University School of Medicine, Washington DC, USA.
- Division of Healthcare Delivery Research, MedStar Health Research Institute, Washington DC, USA.
ORCID-Profile: Erin N Haber
Zitieren
APA
Haber, E. N., Sonti, R., Simkovich, S. M., Pike, C. W., Boxley, C. L., Fong, A., Weintraub, W. S., & Cobb, N. K. (2024). Accuracy of Noninvasive Blood Pressure Monitoring in Critically Ill Adults. Journal of Intensive Care Medicine, 39(7), 665-671. https://doi.org/10.1177/08850666231225173
Vancouver
Haber EN, Sonti R, Simkovich SM, Pike CW, Boxley CL, Fong A, et al. Accuracy of Noninvasive Blood Pressure Monitoring in Critically Ill Adults. Journal of Intensive Care Medicine. 2024;39(7):665-671. doi:10.1177/08850666231225173
BibTeX
@article{Haber2024,
author = {Haber, Erin N and Sonti, Rajiv and Simkovich, Suzanne M and Pike, C William and Boxley, Christian L and Fong, Allan and Weintraub, William S and Cobb, Nathan K},
title = {Accuracy of Noninvasive Blood Pressure Monitoring in Critically Ill Adults},
journal = {Journal of Intensive Care Medicine},
year = {2024},
volume = {39},
number = {7},
pages = {665--671},
doi = {10.1177/08850666231225173}
}
EUVIMED