Ab externo complete glaucoma drainage device occlusion with bulbed polypropylene suture
Digital Journal of Case Reports in Ophthalmology · 2026
DOI: 10.62856/djcro.v10.66 ↗Kurz gefasst
Fachartikel in der Zeitschrift Digital Journal of Case Reports in Ophthalmology, 2026. Verfasst von Daniel Liebman und 2 weiteren Autor:innen.
Abstract (Originalsprache)
We present a novel ab externo glaucoma device occlusion method for patients with hypotony following glaucoma drainage device placement. In this case study, a patient with symptomatic hypotony due to cyclophotocoagulation after glaucoma drainage device implantation underwent occlusion of the drainage device via a modified ab externo approach. After externalizing the tube through a corneal paracentesis incision, a bulbed 3-0 polypropylene suture fragment was inserted into the tube, with subsequent reinsertion of the occluded tube into the anterior chamber. Intraocular pressure increased from 4 mmHg to teens in the first month after the procedure, peaking at 18mmHg at post-operative month four, and stabilizing at 10mmHg on dorzolamide at 14 month follow-up. Visual acuity improved from 20/1000 to 20/400. The technique improves upon previously-described occlusive methods owing to its technical simplicity, safety, minimal invasiveness, and reversibility.
Zitieren
APA
Liebman, D., Nolan, C., & Lin, M. (2026). Ab externo complete glaucoma drainage device occlusion with bulbed polypropylene suture. Digital Journal of Case Reports in Ophthalmology. https://doi.org/10.62856/djcro.v10.66
Vancouver
Liebman D, Nolan C, Lin M. Ab externo complete glaucoma drainage device occlusion with bulbed polypropylene suture. Digital Journal of Case Reports in Ophthalmology. 2026. doi:10.62856/djcro.v10.66
BibTeX
@article{Liebman2026,
author = {Liebman, Daniel and Nolan, Cormac and Lin, Michael},
title = {Ab externo complete glaucoma drainage device occlusion with bulbed polypropylene suture},
journal = {Digital Journal of Case Reports in Ophthalmology},
year = {2026},
doi = {10.62856/djcro.v10.66}
}