Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia

María Agustina Ruggiero, Agustina María Chioli, Milagro Ojeda, Paula María Iacono, Ferrinda Maía, Irina Magalí Forte, Manuela Arca García, Fabián Morales

Revista Bioquímica y Patología Clínica · 16.12.2025 · 90 (1) · 22-26

DOI: 10.62073/sdv51566 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Revista Bioquímica y Patología Clínica, online erschienen am 16.12.2025. Verfasst von María Agustina Ruggiero und 7 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The study of hemostasis includes global screening tests such as activated partial thromboplastin time (APTT) and prothrombin time (PT). APTT and PT can be measured through a wide variety of coagulometers. These coagulometers can be divided into two major groups: those that use optical Clot detection and those that use electromechanical detection.

Objectives

To compare a mechanical method and an optical method to process APTT and PT, and to analyze whether the results reported are equivalent between both methods in a laboratory of a maternal and child hospital. Materials and methods: Blood samples were randomly selected for the analysis of APTT and PT, respectively. Two coagulometers were used: STA Satellite MAX Stago (automated mechanical method) and Coatron X Series TECO (semi-automated optical method).

Results

Based on statistical analysis using Student’s t-test, the results obtained with both types of coagulometer were similar. The interpretation of Pearson correlation values indicated a strong correlation between both methods, with values ranging between 0.5 and 1.0.

Conclusions

The statistical analysis of the data obtained showed that the equipments did not exhibit statistically significant differences in the APTT and PT results; therefore, both methods can be used in the laboratory. The incorporation of the semi-automated Coatron X Series TECO equipment will allow it to be used as a ‘backup’ in the laboratory, both during on-call shifts and in daily routine work.

Zitieren

APA

Ruggiero, M. A., Chioli, A. M., Ojeda, M., Iacono, P. M., Maía, F., Forte, I. M., García, M. A., & Morales, F. (2025). Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia. Revista Bioquímica y Patología Clínica, 90(1), 22-26. https://doi.org/10.62073/sdv51566

Vancouver

Ruggiero MA, Chioli AM, Ojeda M, Iacono PM, Maía F, Forte IM, et al. Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia. Revista Bioquímica y Patología Clínica. 2025;90(1):22-26. doi:10.62073/sdv51566

BibTeX
@article{Ruggiero2025,
  author = {Ruggiero, María Agustina and Chioli, Agustina María and Ojeda, Milagro and Iacono, Paula María and Maía, Ferrinda and Forte, Irina Magalí and García, Manuela Arca and Morales, Fabián},
  title = {Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia},
  journal = {Revista Bioquímica y Patología Clínica},
  year = {2025},
  volume = {90},
  number = {1},
  pages = {22--26},
  doi = {10.62073/sdv51566}
}
Kostenlos starten →