Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia
Revista Bioquímica y Patología Clínica · 16.12.2025 · 90 (1) · 22-26
DOI: 10.62073/sdv51566 ↗Kurz gefasst
Fachartikel in der Zeitschrift Revista Bioquímica y Patología Clínica, online erschienen am 16.12.2025. Verfasst von María Agustina Ruggiero und 7 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
The study of hemostasis includes global screening tests such as activated partial thromboplastin time (APTT) and prothrombin time (PT). APTT and PT can be measured through a wide variety of coagulometers. These coagulometers can be divided into two major groups: those that use optical Clot detection and those that use electromechanical detection.
Objectives
To compare a mechanical method and an optical method to process APTT and PT, and to analyze whether the results reported are equivalent between both methods in a laboratory of a maternal and child hospital. Materials and methods: Blood samples were randomly selected for the analysis of APTT and PT, respectively. Two coagulometers were used: STA Satellite MAX Stago (automated mechanical method) and Coatron X Series TECO (semi-automated optical method).
Results
Based on statistical analysis using Student’s t-test, the results obtained with both types of coagulometer were similar. The interpretation of Pearson correlation values indicated a strong correlation between both methods, with values ranging between 0.5 and 1.0.
Conclusions
The statistical analysis of the data obtained showed that the equipments did not exhibit statistically significant differences in the APTT and PT results; therefore, both methods can be used in the laboratory. The incorporation of the semi-automated Coatron X Series TECO equipment will allow it to be used as a ‘backup’ in the laboratory, both during on-call shifts and in daily routine work.
Zitieren
APA
Ruggiero, M. A., Chioli, A. M., Ojeda, M., Iacono, P. M., Maía, F., Forte, I. M., García, M. A., & Morales, F. (2025). Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia. Revista Bioquímica y Patología Clínica, 90(1), 22-26. https://doi.org/10.62073/sdv51566
Vancouver
Ruggiero MA, Chioli AM, Ojeda M, Iacono PM, Maía F, Forte IM, et al. Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia. Revista Bioquímica y Patología Clínica. 2025;90(1):22-26. doi:10.62073/sdv51566
BibTeX
@article{Ruggiero2025,
author = {Ruggiero, María Agustina and Chioli, Agustina María and Ojeda, Milagro and Iacono, Paula María and Maía, Ferrinda and Forte, Irina Magalí and García, Manuela Arca and Morales, Fabián},
title = {Estudio comparativo entre un método mecánico automatizado y un método óptico semiautomatizado para la medición de las pruebas básicas del área de hemostasia},
journal = {Revista Bioquímica y Patología Clínica},
year = {2025},
volume = {90},
number = {1},
pages = {22--26},
doi = {10.62073/sdv51566}
}
EUVIMED