EUVIMEDEUVIMED entdecken
Crossref

In Vitro Accuracy Assessment of a Surgical Navigation System for Dental Implantology

Chang-Kai Chen, Da‐Yo Yuh, Cheng-Feng Tsai, Po‐Jan Kuo, Cheng‐Yang Chiang

臺灣牙周病醫學會雜誌 · 01.12.2015 · 20 (4) · 276-287

DOI: 10.6121/tap.2015.20.4.03 ↗

Kurz gefasst

Fachartikel in der Zeitschrift 臺灣牙周病醫學會雜誌, erschienen am 01.12.2015. Verfasst von Chang-Kai Chen und 4 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

During the process of dental implantation, mistaken use of a surgical drill may cause damage to important anatomical structures. As a result, the use of computer-assisted systems to prevent damage has become commonplace among many dentists. Therefore, it is necessary to be aware of the accuracy of the surgical navigation system, and system accuracy is also an important reference for error range used by dentists in clinical practice. Materials and Methods: This study included the carrying out of an accuracy test on surgical navigation system and freehand drilling under an in vitro environment. Twelve dental implant plans were separately performed in two models; in addition, a surgical navigation system and freehand drilling were used in a simulated dental implant surgery. Furthermore, the accuracy of the planned and actual positions of the implants was assessed.

Results

The entry lateral error of the surgical navigation system and freehand drilling was 0.73 ± 0.45 mm and 1.18 ± 0.25 mm, respectively; the apex lateral error was 0.71 ± 0.51 mm and 1.12 ± 0.49 mm, respectively; the apex longitudinal error always 0.25 ± 0.17 mm and 0.63 ± 0.43 mm, respectively; the total error was 0.78 ± 0.48 mm and 1.36 ± 0.44 mm, respectively; and the angular error was 2.09 ± 1.03 degrees and 8.58 ± 3.57 degrees, respectively. The results indicate that the error while using the surgical navigation system was smaller than the error of freehand drilling.

Conclusion

This study included the assessment of drilling accuracy of the surgical navigation system in a gypsum model which revealed that the surgical navigation system was more accurate than freehand drilling; however, there are significant differences between the gypsum model and actual patients. Therefore, future clinical studies are warranted in order to provide more valuable data for reference.

Schlagwörter

surgical navigation systemsdental implant surgerydrilling accuracycomputer-assisted implant surgerygypsum models

Zitieren

APA

Chen, C., Yuh, D., Tsai, C., Kuo, P., & Chiang, C. (2015). In Vitro Accuracy Assessment of a Surgical Navigation System for Dental Implantology. 臺灣牙周病醫學會雜誌, 20(4), 276-287. https://doi.org/10.6121/tap.2015.20.4.03

Vancouver

Chen C, Yuh D, Tsai C, Kuo P, Chiang C. In Vitro Accuracy Assessment of a Surgical Navigation System for Dental Implantology. 臺灣牙周病醫學會雜誌. 2015;20(4):276-287. doi:10.6121/tap.2015.20.4.03

BibTeX
@article{Chen2015,
  author = {Chen, Chang-Kai and Yuh, Da‐Yo and Tsai, Cheng-Feng and Kuo, Po‐Jan and Chiang, Cheng‐Yang},
  title = {In Vitro Accuracy Assessment of a Surgical Navigation System for Dental Implantology},
  journal = {臺灣牙周病醫學會雜誌},
  year = {2015},
  volume = {20},
  number = {4},
  pages = {276--287},
  doi = {10.6121/tap.2015.20.4.03}
}
Kostenlos starten →