Accuracy of immediate dental implant placement with task‐autonomous robotic system and navigation system: An in vitro study
Clinical Oral Implants Research · 2024-08 · 35 (8) · 973-983
DOI: 10.1111/clr.14104 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, online erschienen am 29.05.2023. Verfasst von Jinyan Chen und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Second Dental Center, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. Hauptthemen laut MEDLINE: Robotic Surgical Procedures. Die Arbeit wurde bisher 49-mal zitiert (Crossref) und nennt 41 Literaturangaben. Gefördert von: Clinical Research Plan of SHDC, CAMS Innovation Fund for Medical Sciences (CIFMS), Research Discipline Fund from Shanghai Ninth People's Hospital.
Abstract (Originalsprache)
Objectives
The aim of this study was to compare the accuracy of dental implant placement in a single tooth gap, including the postextraction site and healed site, using a task‐autonomous robotic system and a dynamic navigation system. Materials and Methods Forty partially edentulous models requiring both immediate and conventional implant placement were randomly divided into a robotic system group and a navigation system group. The coronal, apical, and angular deviations of the implants were measured and assessed between the groups.
Results
The deviations in immediate implant placement were compared between the robotic system and dynamic navigation system groups, showing a mean (±SD) coronal deviation of 0.86 ± 0.36 versus 0.70 ± 0.21 mm ( p = .101), a mean apical deviation of 0.77 ± 0.34 versus 0.95 ± 0.38 mm ( p = .127), and a mean angular deviation of 1.94 ± 0.66° versus 3.44 ± 1.38° ( p < .001). At the healed site, significantly smaller coronal deviation (0.46 ± 0.29 vs. 0.70 ± 0.30 mm, p = .005), apical deviation (0.56 ± 0.30 vs. 0.85 ± 0.25 mm, p < .001), and angular deviation (1.36 ± 0.54 vs. 1.80 ± 0.70 mm, p = .034) were found in the robotic system group than in the dynamic navigation group.
Conclusions
The position in both immediate and conventional implant placement was more precise with the task‐autonomous robotic system than with the dynamic navigation system. Its performance in actual clinical applications should be confirmed in further trials.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Second Dental Center, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
- College of Stomatology, Shanghai Jiao Tong University, Shanghai, China.
- National Center for Stomatology, Shanghai, China.
- National Clinical Research Center for Oral Diseases, Shanghai, China.
- Shanghai Key Laboratory of Stomatology, Shanghai, China.
- Shanghai Research Institute of Stomatology, Shanghai, China.
ORCID-Profile: Jinyan Chen
Förderung
- Clinical Research Plan of SHDC
- CAMS Innovation Fund for Medical Sciences (CIFMS)
- Research Discipline Fund from Shanghai Ninth People's Hospital
- Huangpu District Industrial Support Fund
- ITI Research Grant
Zitieren
APA
Chen, J., Zhuang, M., Tao, B., Wu, Y., Ye, L., & Wang, F. (2024). Accuracy of immediate dental implant placement with task‐autonomous robotic system and navigation system: An in vitro study. Clinical Oral Implants Research, 35(8), 973-983. https://doi.org/10.1111/clr.14104
Vancouver
Chen J, Zhuang M, Tao B, Wu Y, Ye L, Wang F. Accuracy of immediate dental implant placement with task‐autonomous robotic system and navigation system: An in vitro study. Clinical Oral Implants Research. 2024;35(8):973-983. doi:10.1111/clr.14104
BibTeX
@article{Chen2024,
author = {Chen, Jinyan and Zhuang, Minjie and Tao, Baoxin and Wu, Yiqun and Ye, Lijuan and Wang, Feng},
title = {Accuracy of immediate dental implant placement with task‐autonomous robotic system and navigation system: An in vitro study},
journal = {Clinical Oral Implants Research},
year = {2024},
volume = {35},
number = {8},
pages = {973--983},
doi = {10.1111/clr.14104}
}