Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique

Olivia Ganna, Akhirida Putri, Wenda Anastasia, Suherman Hardi Saputro, Mira Maya Kumala, Nurwahyu, Purwanto

Journal Interdisciplinary Health · 12.02.2026 · 2 (1) · 45-54

DOI: 10.61099/jih.v2i1.301 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal Interdisciplinary Health, online erschienen am 12.02.2026. Verfasst von Olivia Ganna und 6 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Computed Tomography Angiography (CTA) of the cerebral carotid arteries is a key non-invasive modality for evaluating vascular abnormalities. Image quality and radiation dose are strongly influenced by tube voltage (kVp). Lower tube voltage may enhance iodine attenuation and reduce radiation exposure. This study aimed to compare vascular image quality and radiation dose between low tube voltage (70 kVp) and standard tube voltage (100 kVp) in cerebral carotid CTA using bolus tracking. Research Methodology: A quantitative comparative retrospective study was conducted on 20 patients, divided into two groups: 70 kVp (n=10) and 100 kVp (n=10). Image quality was assessed using arterial enhancement (Hounsfield Unit, HU) and Signal-to-Noise Ratio (SNR) measured at the ascending aorta, common carotid artery, and middle cerebral artery using Region of Interest (ROI) analysis. Radiation dose parameters included CTDIvol and Dose Length Product (DLP). Data normality was tested using the Shapiro–Wilk, followed by the the Mann–Whitney and independent t-test for comparative analysis.

Results

The 70 kVp protocol significantly increased arterial enhancement (e.g., ascending aorta: 697.36±81.06 HU vs 328.25±34.75 HU) and SNR (29.43±16.29 vs 15.68±7.95) compared to 100 kVp (p < 0.05). Radiation dose was substantially reduced, with CTDIvol decreasing from 8.08 mGy to 2.82 mGy and DLP from 284.66±10.92 mGy·cm to 102.12±3.08 mGy·cm (p < 0.05).

Conclusion

Low tube voltage (70 kVp) significantly improves vascular image quality while reducing radiation dose in cerebral carotid CTA. This protocol supports radiation optimization principles and can be considered an effective alternative in clinical practice.

Zitieren

APA

Ganna, O., Putri, A., Anastasia, W., Saputro, S. H., Kumala, M. M., Nurwahyu, & Purwanto (2026). Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique. Journal Interdisciplinary Health, 2(1), 45-54. https://doi.org/10.61099/jih.v2i1.301

Vancouver

Ganna O, Putri A, Anastasia W, Saputro SH, Kumala MM, Nurwahyu, et al. Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique. Journal Interdisciplinary Health. 2026;2(1):45-54. doi:10.61099/jih.v2i1.301

BibTeX
@article{Ganna2026,
  author = {Ganna, Olivia and Putri, Akhirida and Anastasia, Wenda and Saputro, Suherman Hardi and Kumala, Mira Maya and Nurwahyu and Purwanto},
  title = {Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique},
  journal = {Journal Interdisciplinary Health},
  year = {2026},
  volume = {2},
  number = {1},
  pages = {45--54},
  doi = {10.61099/jih.v2i1.301}
}
Kostenlos starten →