Fundamental Study on Fluoroscopy Dose Rate Reduction in Pediatric Angiography Using an X-ray Cut Filter

Yusuke Takada, Hidemitsu Otsu, Naoki Yoshikawa, Kojiro Nishijima

Japanese Journal of Radiological Technology · 2026 · 82 (7) · n/a

DOI: 10.6009/jjrt.26-1617 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2026. Verfasst von Yusuke Takada und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Oita Prefectural Hospital. Hauptthemen laut MEDLINE: Angiography, Radiation Dosage. Die Arbeit nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

We evaluated the ability of an MJP X-ray cut filter (hereafter referred to as “cut filter”) to reduce fluoroscopic dose rate and assessed its impact on image quality in pediatric angiography.

Methods

Fundamental assessments included fluoroscopic dose rate measurements, spectral analysis, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and visual evaluation were conducted. Body thickness was also measured in 412 pediatric patients aged 10 years or younger. Image visibility was visually assessed using a paired-comparison method.

Results

The use of a cut filter reduced the fluoroscopic dose rate by 15%–25%. Spectral analysis confirmed that the low-energy component was effectively reduced via use of a cut filter. Changes in SNR ranged from −1.7% to +3.0%. CNR decreased by 27% when polymethyl methacrylate (PMMA) thickness was 20 cm. Visual assessment showed an average degree of preference of 0.119 at a PMMA thickness of 20 cm. Respective average body thicknesses at 0 and 10 years of age were 10.5±0.63 and 17.0±2.04 cm, respectively.

Conclusion

An MJP X-ray cut filter can reduce fluoroscopic dose rate while maintaining sufficient image visibility in pediatric angiography, particularly for patients with a body thickness of 15 cm or less.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AngiographyRadiation DosageHumansFluoroscopyChildInfantChild, Preschool

Stichwörter der Autor:innen

X-ray cut filterangiographydose reductionpediatric

Autor:innen und Institutionen

  • Department of Radiology, Oita Prefectural Hospital.

Zitieren

APA

Takada, Y., Otsu, H., Yoshikawa, N., & Nishijima, K. (2026). Fundamental Study on Fluoroscopy Dose Rate Reduction in Pediatric Angiography Using an X-ray Cut Filter. Japanese Journal of Radiological Technology, 82(7), n/a. https://doi.org/10.6009/jjrt.26-1617

Vancouver

Takada Y, Otsu H, Yoshikawa N, Nishijima K. Fundamental Study on Fluoroscopy Dose Rate Reduction in Pediatric Angiography Using an X-ray Cut Filter. Japanese Journal of Radiological Technology. 2026;82(7):n/a. doi:10.6009/jjrt.26-1617

BibTeX
@article{Takada2026,
  author = {Takada, Yusuke and Otsu, Hidemitsu and Yoshikawa, Naoki and Nishijima, Kojiro},
  title = {Fundamental Study on Fluoroscopy Dose Rate Reduction in Pediatric Angiography Using an X-ray Cut Filter},
  journal = {Japanese Journal of Radiological Technology},
  year = {2026},
  volume = {82},
  number = {7},
  pages = {n/a},
  doi = {10.6009/jjrt.26-1617}
}
Kostenlos starten →