Evaluation of Imaging Uniformity and Optimal Positioning of Head Coils from Different Manufacturers Using Vacuum Immobilization in Low-birth-weight Infants
Japanese Journal of Radiological Technology · 2026
DOI: 10.6009/jjrt.26-1582 ↗Kurz gefasst
Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2026. Verfasst von Hiroya Asou und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Shimane University Hospital. Hauptthemen laut MEDLINE: Infant, Low Birth Weight, Head, Immobilization, Magnetic Resonance Imaging. Die Arbeit nennt 8 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To investigate the optimal positioning for low-birth-weight infants using a vacuum fixation device, considering the most suitable position for each manufacturer's coil. METHODS: Image data of low-birth-weight infants scanned at the institution were measured to assess their physical characteristics. Uniformity was evaluated by imaging a common phantom under identical imaging conditions using the head coils provided by GE Healthcare (Milwaukee, WI, USA), Siemens Healthineers (Erlangen, Germany), and Philips Healthcare (Best, the Netherlands). RESULTS: The physical characteristics of low-birth-weight infants allowed them to fit entirely within all manufacturers' coils. Uniformity varied depending on the coil type and the position within the coil: uniformity was highest in the deeper position for the GE 8-channel coil, in the central position for the Siemens 20-channel coil, in the deeper position for the Philips 32-channel coil, and in the central position for the Philips 20-channel coil. CONCLUSION: When using a vacuum fixation device in low-birth-weight infants, optimal positioning was achieved by adjusting for the thickness of the fixation device and placing the infant at the position within the coil where imaging uniformity was highest.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Shimane University Hospital.
- Department of Radiology, Faculty of Medicine, Shimane University.
Zitieren
APA
Asou, H., Matsuo, K., Miyahara, Y., Katsube, T., Yoshizako, T., & Kaji, Y. (2026). Evaluation of Imaging Uniformity and Optimal Positioning of Head Coils from Different Manufacturers Using Vacuum Immobilization in Low-birth-weight Infants. Japanese Journal of Radiological Technology. https://doi.org/10.6009/jjrt.26-1582
Vancouver
Asou H, Matsuo K, Miyahara Y, Katsube T, Yoshizako T, Kaji Y. Evaluation of Imaging Uniformity and Optimal Positioning of Head Coils from Different Manufacturers Using Vacuum Immobilization in Low-birth-weight Infants. Japanese Journal of Radiological Technology. 2026. doi:10.6009/jjrt.26-1582
BibTeX
@article{Asou2026,
author = {Asou, Hiroya and Matsuo, Kazuaki and Miyahara, Yoshinori and Katsube, Takashi and Yoshizako, Takeshi and Kaji, Yasushi},
title = {Evaluation of Imaging Uniformity and Optimal Positioning of Head Coils from Different Manufacturers Using Vacuum Immobilization in Low-birth-weight Infants},
journal = {Japanese Journal of Radiological Technology},
year = {2026},
doi = {10.6009/jjrt.26-1582}
}
EUVIMED