Influence of Calculated Equivalent Doses and Effective Dose due to Differences in Software Version on Radiation Dose Management System

Yasutaka Takei, Ryo Moriwake, Noriaki Akagi

Japanese Journal of Radiological Technology · 2022 · 78 (11) · 1323-1332

DOI: 10.6009/jjrt.2022-1312 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2022. Verfasst von Yasutaka Takei und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare. Hauptthemen laut MEDLINE: Algorithms, Liver. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 9 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To verify the influence of the software version of the dose control system on the equivalent dose and effective dose calculation values. METHODS: We performed chest CT and liver to pelvic CT imaging with a human phantom placed on a CT bed. After the imaging was completed, the radiation dose structured reports (RDSRs) generated by the equipment were transferred to several dose management systems with different software versions for equivalent dose and effective dose calculations. RESULTS: The equivalent and effective doses calculated from the same RDSR differed depending on the software version of the dose management system with a difference of approximately 2 times for the effective dose and up to 50 times for the equivalent dose. CONCLUSION: It is considered that the voxel phantom geometry and dose calculation algorithm may differ depending on the software version of the dose management system. Careful attention should be paid in handling the dose calculation values because the exposure explanations and risk assessments using the equivalent doses, and effective dose calculated by the dose management systems may be overestimated.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsLiverHumansPhantoms, ImagingRadiation DosageSoftware

Stichwörter der Autor:innen

dose management systemeffective doseequivalent dose

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare.
  • Department of Radiology, Kawasaki Medical School Hospital.
  • Department of Medical Support, Division of Radiological Technology, Okayama University Hospital.

Zitieren

APA

Takei, Y., Moriwake, R., & Akagi, N. (2022). Influence of Calculated Equivalent Doses and Effective Dose due to Differences in Software Version on Radiation Dose Management System. Japanese Journal of Radiological Technology, 78(11), 1323-1332. https://doi.org/10.6009/jjrt.2022-1312

Vancouver

Takei Y, Moriwake R, Akagi N. Influence of Calculated Equivalent Doses and Effective Dose due to Differences in Software Version on Radiation Dose Management System. Japanese Journal of Radiological Technology. 2022;78(11):1323-1332. doi:10.6009/jjrt.2022-1312

BibTeX
@article{Takei2022,
  author = {Takei, Yasutaka and Moriwake, Ryo and Akagi, Noriaki},
  title = {Influence of Calculated Equivalent Doses and Effective Dose due to Differences in Software Version on Radiation Dose Management System},
  journal = {Japanese Journal of Radiological Technology},
  year = {2022},
  volume = {78},
  number = {11},
  pages = {1323--1332},
  doi = {10.6009/jjrt.2022-1312}
}
Kostenlos starten →