Evaluation and Examination of Accuracy for the Conversion Factors of Effective Dose per Dose^|^#8211;Length Product

Masanao Kobayashi, Tomoko Ootsuka, Syoichi Suzuki

Japanese Journal of Radiological Technology · 2013 · 69 (1) · 19-27

DOI: 10.6009/jjrt.2013_jsrt_69.1.19 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2013. Verfasst von Masanao Kobayashi und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Fujita Health University Hospital. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 16 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The purpose of this paper is to reappraise the accuracy of a conversion coefficient (k) reported by International Commission on Radiological Protection Publication 102 Table A.2. The effective doses of the routine head computed tomography (CT), the routine chest CT, the perfusion CT, and the coronary CT were evaluated using the conversion coefficient (adult head: 0.021 mSv·mGy⁻¹·cm⁻¹, adult chest: 0.014 mSv·mGy⁻¹·cm⁻¹). The dose length product (DLP) used the value displayed on the console on each scanning condition. The effective doses were evaluated using a human body type phantom (Alderson Rando phantom) and thermoluminescent dosimeter (TLD) elements for comparison with the converted value. This paper reported that the effective doses evaluated from conversion coefficient became different by 0.3 mSv (17%) compared with measurements, the effective dose computed with the conversion coefficient of the adult chest may be underestimated by 45%, and the bolus-tracking which scans the narrow beams should not use a conversion coefficient.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageTomography, X-Ray ComputedAdultEvaluation Studies as TopicHumansPhantoms, ImagingRadiation Protection

Autor:innen und Institutionen

  • Department of Radiology, Fujita Health University Hospital.

Zitieren

APA

Kobayashi, M., Ootsuka, T., & Suzuki, S. (2013). Evaluation and Examination of Accuracy for the Conversion Factors of Effective Dose per Dose^|^#8211;Length Product. Japanese Journal of Radiological Technology, 69(1), 19-27. https://doi.org/10.6009/jjrt.2013_jsrt_69.1.19

Vancouver

Kobayashi M, Ootsuka T, Suzuki S. Evaluation and Examination of Accuracy for the Conversion Factors of Effective Dose per Dose^|^#8211;Length Product. Japanese Journal of Radiological Technology. 2013;69(1):19-27. doi:10.6009/jjrt.2013_jsrt_69.1.19

BibTeX
@article{Kobayashi2013,
  author = {Kobayashi, Masanao and Ootsuka, Tomoko and Suzuki, Syoichi},
  title = {Evaluation and Examination of Accuracy for the Conversion Factors of Effective Dose per Dose^|^#8211;Length Product},
  journal = {Japanese Journal of Radiological Technology},
  year = {2013},
  volume = {69},
  number = {1},
  pages = {19--27},
  doi = {10.6009/jjrt.2013_jsrt_69.1.19}
}