Design of a Microcontroller Based and X-Ray Waveform Independent kVp-Meter
American Journal of Biomedical Engineering · 2012
DOI: 10.5923/j.ajbe.20110101.07 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Biomedical Engineering, 2012. Verfasst von Y. Ülgen und 1 weiteren Autor:innen.
Abstract (Originalsprache)
The kVp setting is one of the major factors affecting the image quality in X-ray imaging and should be annually measured and calibrated if necessary. In this work, a kVp-meter is designed around the ATmega16 (Atmel) microcontroller, based on the physical principle that the linear attenuation coefficient of materials, namely copper has a smooth dependence on the energy level of the X-ray photons. Based on the logarithm of the ratio of the radiation intensities through 0.5mm and 1mm thick copper filters, a look-up table is generated in the range 60-120kVp. Logarithmic operation increased the precision at higher kVp values. Since sampling is performed over the exposure period in a continuous manner, the measurement is not affected by the X-ray waveform. A prototype unit was built and the performance was tested in terms of accuracy, precision and reliability.
Zitieren
APA
Ülgen, Y., & Tümer, M. (2012). Design of a Microcontroller Based and X-Ray Waveform Independent kVp-Meter. American Journal of Biomedical Engineering. https://doi.org/10.5923/j.ajbe.20110101.07
Vancouver
Ülgen Y, Tümer M. Design of a Microcontroller Based and X-Ray Waveform Independent kVp-Meter. American Journal of Biomedical Engineering. 2012. doi:10.5923/j.ajbe.20110101.07
BibTeX
@article{lgen2012,
author = {Ülgen, Y. and Tümer, M.},
title = {Design of a Microcontroller Based and X-Ray Waveform Independent kVp-Meter},
journal = {American Journal of Biomedical Engineering},
year = {2012},
doi = {10.5923/j.ajbe.20110101.07}
}
EUVIMED