Phantom-Based Comparison of Image Quality and Average Glandular Dose in Digital Mammography Systems

Mubarak Yosif, Murtada Ahmed

International Journal of Applied Science and Research · 2026 · 09 (01) · 162-170

DOI: 10.56293/ijasr.2026.6918 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Applied Science and Research, 2026. Verfasst von Mubarak Yosif und 1 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluates and compares image quality and average glandular dose (AGD) across six digital mammography systems from Siemens, GE, and Hologic, using standardized phantom testing in a clinical environment.

Methods

Images were acquired with the Pro-MAM Accreditation FF phantom and PMMA slabs (2–6 cm) to simulate different breast thicknesses. All systems operated in Auto Filter mode. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured from phantom images. AGD, kVp, mAs, and ACR scores were recorded.

Results

All systems passed ACR phantom image quality criteria. GE Prestina demonstrated the highest SNR values across thicknesses with efficient dose performance. Siemens systems showed consistent balanced results. Hologic systems performed adequately but limited access to raw data restricted detailed analysis. AGD values varied significantly between systems but were within clinical limits.

Conclusions

All systems achieved acceptable image quality and dose levels according to clinical standards. However, significant variability in SNR and AGD performance was observed, highlighting differences in optimization among vendors.

Zitieren

APA

Yosif, M., & Ahmed, M. (2026). Phantom-Based Comparison of Image Quality and Average Glandular Dose in Digital Mammography Systems. International Journal of Applied Science and Research, 09(01), 162-170. https://doi.org/10.56293/ijasr.2026.6918

Vancouver

Yosif M, Ahmed M. Phantom-Based Comparison of Image Quality and Average Glandular Dose in Digital Mammography Systems. International Journal of Applied Science and Research. 2026;09(01):162-170. doi:10.56293/ijasr.2026.6918

BibTeX
@article{Yosif2026,
  author = {Yosif, Mubarak and Ahmed, Murtada},
  title = {Phantom-Based Comparison of Image Quality and Average Glandular Dose in Digital Mammography Systems},
  journal = {International Journal of Applied Science and Research},
  year = {2026},
  volume = {09},
  number = {01},
  pages = {162--170},
  doi = {10.56293/ijasr.2026.6918}
}
Kostenlos starten →