Average Glandular Dose (AGD) and Radiation Dose Optimization in Screen-Film and Digital X-ray Mammography
Applied Sciences · 31.10.2023 · 13 (21) · 11901
DOI: 10.3390/app132111901 ↗Kurz gefasst
Fachartikel in der Zeitschrift Applied Sciences, online erschienen am 31.10.2023. Verfasst von Ibrahim I. Suliman und 8 weiteren Autor:innen. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Imam Mohammad Ibn Saud Islamic University (IMSIU).
Abstract (Originalsprache)
We determined the average glandular dose (AGD) from the craniocaudally (CC) and mediolateral oblique (MLO) views of 496 breasts (247 women) at eight clinics in Sudan. The incident air kerma from the X-ray tube output values and the typical patient-specific breast exposure factors were measured. The AGD values were inferred from the measured incident air kerma and breast-specific dose-conversion coefficients. The AGD per CC and MLO projection and per woman ranged from 0.56 to 2.89 mGy (average: 1.36), 0.48 to 2.08 mGy (average: 1.19), and 2.08 to 9.94 mGy (average: 5.10). The proposed national diagnostic reference levels (mGy) for digital mammography are 1.8 and 1.6 mGy for CC and MLO projection, respectively. Establishing the proposed diagnostic reference levels is an essential step in ensuring patient protection from radiation and will help promote dose optimization for X-ray mammography at the national level and beyond. These results provide important baseline data that can be used to formulate national diagnostic reference levels.
Förderung
- Imam Mohammad Ibn Saud Islamic University (IMSIU)
Zitieren
APA
Suliman, I. I., Mohamed, S., Mahadi, A., Bashier, E., Farah, A., Hassan, N., Ibrahem, M. A., Ali, M. K. M., & Ahmed, N. A. (2023). Average Glandular Dose (AGD) and Radiation Dose Optimization in Screen-Film and Digital X-ray Mammography. Applied Sciences, 13(21), 11901. https://doi.org/10.3390/app132111901
Vancouver
Suliman II, Mohamed S, Mahadi A, Bashier E, Farah A, Hassan N, et al. Average Glandular Dose (AGD) and Radiation Dose Optimization in Screen-Film and Digital X-ray Mammography. Applied Sciences. 2023;13(21):11901. doi:10.3390/app132111901
BibTeX
@article{Suliman2023,
author = {Suliman, Ibrahim I. and Mohamed, Sawsan and Mahadi, Alaa and Bashier, Einas and Farah, Atifa and Hassan, Nada and Ibrahem, Moez A. and Ali, Mohammed Khalil Mohammed and Ahmed, Nada A.},
title = {Average Glandular Dose (AGD) and Radiation Dose Optimization in Screen-Film and Digital X-ray Mammography},
journal = {Applied Sciences},
year = {2023},
volume = {13},
number = {21},
pages = {11901},
doi = {10.3390/app132111901}
}
EUVIMED