Feasibility Study of Using Detector-Scattered Photons for Attenuation Correction in CT-less PET Imaging

Satyam Tiwari, Sushil Sharma, Paweł Moskal

Bio-Algorithms and Med-Systems · 16.12.2025 · 21 (Special issue (New Trends)) · 13-20

DOI: 10.5604/01.3001.0055.4655 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bio-Algorithms and Med-Systems, erschienen am 16.12.2025. Verfasst von Satyam Tiwari und 2 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Quantitative PET requires accurate attenuation correction (AC), which is conventionally achieved using CT-derived maps, incurring extra dose and complexity. This study explores a CT-less AC approach for the plastic scintillator-based J-PET scanner by investigating whether detector-scattered photons (DS-LORs) can serve as an intrinsic transmission source.



Methods: • Simulation Setup: The study utilized GATE v9.1 software to simulate the modular J-PET scanner, which consists of plastic scintillator strips where 511 keV photons predominantly undergo Compton scattering. Two configurations were evaluated: a uniform water cylinder and a NEMA IQlike phantom containing a cold lung insert and six hot spheres to simulate heterogeneous activity. • Analysis & reconstruction: The study isolated triple-hit coincidences to analyze angular deviations. DS-LORs were identified using a specific scatter test metric based on interaction positions and detection times. Attenuation maps were reconstructed using Filtered Back Projection (FBP) applied to the ratio of water-to-air sinograms.

Results

Due to the properties of plastic scintillators, DS-LORs constituted approximately 50% of all detected prompts, providing a substantial dataset for analysis. In a simulated 10-minute acquisition, the noise propagation analysis yielded a Signal-to-Noise Ratio (SNR) of ≈ 33.9 for water. This is significantly higher than conventional crystal-based PET, which yields only ≈ 5% DS-LORs and a lower SNR. Reconstructions successfully resolved the NEMA phantom features, including the cold lung insert and hot spheres, particularly when phantom-scattered events were rejected.



Conclusions and future directions: This study confirms that DS-LORs retain sufficient attenuation information to estimate μ-maps, demonstrating the potential for fully CT-less PET imaging. While the current implementation faces limitations like reliance on blank scans and activity-attenuation crosstalk, the high statistics of DS-LORs are a foundation for developing joint reconstruction algorithms aiming towards clinical application.

Zitieren

APA

Tiwari, S., Sharma, S., & Moskal, P. (2025). Feasibility Study of Using Detector-Scattered Photons for Attenuation Correction in CT-less PET Imaging. Bio-Algorithms and Med-Systems, 21(Special issue (New Trends)), 13-20. https://doi.org/10.5604/01.3001.0055.4655

Vancouver

Tiwari S, Sharma S, Moskal P. Feasibility Study of Using Detector-Scattered Photons for Attenuation Correction in CT-less PET Imaging. Bio-Algorithms and Med-Systems. 2025;21(Special issue (New Trends)):13-20. doi:10.5604/01.3001.0055.4655

BibTeX
@article{Tiwari2025,
  author = {Tiwari, Satyam and Sharma, Sushil and Moskal, Paweł},
  title = {Feasibility Study of Using Detector-Scattered Photons for Attenuation Correction in CT-less PET Imaging},
  journal = {Bio-Algorithms and Med-Systems},
  year = {2025},
  volume = {21},
  number = {Special issue (New Trends)},
  pages = {13--20},
  doi = {10.5604/01.3001.0055.4655}
}
Kostenlos starten →