Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans

Rafel Tappouni, Bradley Mathers

ISRN Radiology · 2013

DOI: 10.5402/2013/457396 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ISRN Radiology, 2013. Verfasst von Rafel Tappouni und 1 weiteren Autor:innen. Beteiligte Einrichtung: Radiology Department, Penn State Hershey Medical Center, 500 University Drive, Hershey, PA 17033, USA. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare the effectiveness of the bismuth breast shield and partial CT scan in reducing entrance skin dose and to evaluate the effect of the breast shield on image quality (IQ).

Methods

Nanodots were placed on an adult anthropomorphic phantom. Standard chest CT, CT with shield, and partial CT were performed. Nanodot readings and effective doses were recorded. 50 patients with chest CTs obtained both with and without breast shields were reviewed. IQ was evaluated by two radiologists and by measuring Hounsfield units (HUs) and standard deviation (SD) of HU in anterior subcutaneous region.

Results

Breast shield and the partial CT scans reduced radiation to the anterior chest by 38% and 16%, respectively. Partial CT increased dose to the posterior chest by 37% and effective dose by 8%. Change in IQ in shield CT was observed in the anterior chest wall. Significant change in IQ was observed in 5/50 cases. The shield caused an increase of 20 HU () and a 1.86 reduction in SD of HU () in the anterior compared to posterior subcutaneous regions. Summary . Bismuth breast shield is more effective than the partial CT in reducing entrance skin dose while maintaining image quality.

Autor:innen und Institutionen

  • Radiology Department, Penn State Hershey Medical Center, 500 University Drive, Hershey, PA 17033, USA.
  • Internal Medicine Department, Penn State Hershey Medical Center, 500 University Drive, Hershey, PA 17033, USA.

Zitieren

APA

Tappouni, R., & Mathers, B. (2013). Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans. ISRN Radiology. https://doi.org/10.5402/2013/457396

Vancouver

Tappouni R, Mathers B. Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans. ISRN Radiology. 2013. doi:10.5402/2013/457396

BibTeX
@article{Tappouni2013,
  author = {Tappouni, Rafel and Mathers, Bradley},
  title = {Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans},
  journal = {ISRN Radiology},
  year = {2013},
  doi = {10.5402/2013/457396}
}
Kostenlos starten →