Dosimetric comparison between microSelectron iridium-192 and flexi cobalt-60 sources in high-dose-rate brachytherapy using Geant4 Monte Carlo code

Tsige Yhidego Gebremariam, Ghazale Geraily, Hussam Hameed Jassim, Somayeh Gholami

Journal of Contemporary Brachytherapy · 2023 · 15 (2) · 141-147

DOI: 10.5114/jcb.2023.126238 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Contemporary Brachytherapy, 2023. Verfasst von Tsige Yhidego Gebremariam und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran. Die Arbeit wurde bisher 6-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Manufacturing of miniaturized high activity iridium-192 (192Ir) sources have been made a market preference in modern brachytherapy. Smaller dimensions of the sources are flexible for smaller diameter of the applicators, and it is also suitable for interstitial implants. Presently, cobalt-60 (60Co) sources have been commercialized as an alternative to 192Ir sources for high-dose-rate (HDR) brachytherapy, since 60Co source have an advantage of longer half-life comparing with 192Ir source. One of them is the HDR 60Co Flexisource manufactured by Elekta. The purpose of this study was to compare the TG-43 dosimetric parameters of HDR flexi 60Co and HDR microSelectron 192Ir sources. Material and methods: Monte Carlo simulation code of Geant4 (v.11.0) was applied. Following the recommendations of AAPM TG-43 formalism report, Monte Carlo code of HDR flexi 60Co and HDR microSelectron 192Ir was validated by calculating radial dose function, anisotropy function, and dose-rate constants in a water phantom. Finally, results of both radionuclide sources were compared.

Results

The calculated dose-rate constants per unit air-kerma strength in water medium were 1.108 cGy h-1U-1 for HDR microSelectron 192Ir, and 1.097 cGy h-1U-1 for HDR flexi 60Co source, with the percentage uncertainty of 1.1% and 0.2%, respectively. The values of radial dose function for distances above 22 cm for HDR flexi 60Co source were higher than that of the other source. The anisotropic values sharply increased to the longitudinal sides of HDR flexi 60Co source, and the rise was comparatively sharper to that of the other source.

Conclusions

The primary photons from the lower-energy HDR microSelectron 192Ir source have a limited range and are partially attenuated when considering the results of radial and anisotropic dose distribution functions. This implies that a HDR flexi 60Co radionuclide could be used to treat tumors beyond the source compared with a HDR microSelectron 192Ir source, despite the fact that 192Ir has a lower exit dose than HDR flexi 60Co radionuclide source.

Schlagwörter

Stichwörter der Autor:innen

Geant4Monte Carlo dosimetrybrachytherapyflexi HDR 60ComicroSelectron HDR 192Ir

Autor:innen und Institutionen

  • Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Department of Medical Physics and Biomedical Engineering, School of Medicine, Tehran University of Medical Sciences, International Campus, Tehran, Iran.
  • Department of Physics, College of Natural Sciences, Aksum University, Axum, Ethiopia.
  • Radiation Oncology Research Center, Cancer Institute, Tehran University of Medical Science, Tehran, Iran.
  • School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
  • Najaf Cancer Center, Najaf Hospital, Najaf, Iraq.

Zitieren

APA

Gebremariam, T. Y., Geraily, G., Jassim, H. H., & Gholami, S. (2023). Dosimetric comparison between microSelectron iridium-192 and flexi cobalt-60 sources in high-dose-rate brachytherapy using Geant4 Monte Carlo code. Journal of Contemporary Brachytherapy, 15(2), 141-147. https://doi.org/10.5114/jcb.2023.126238

Vancouver

Gebremariam TY, Geraily G, Jassim HH, Gholami S. Dosimetric comparison between microSelectron iridium-192 and flexi cobalt-60 sources in high-dose-rate brachytherapy using Geant4 Monte Carlo code. Journal of Contemporary Brachytherapy. 2023;15(2):141-147. doi:10.5114/jcb.2023.126238

BibTeX
@article{Gebremariam2023,
  author = {Gebremariam, Tsige Yhidego and Geraily, Ghazale and Jassim, Hussam Hameed and Gholami, Somayeh},
  title = {Dosimetric comparison between microSelectron iridium-192 and flexi cobalt-60 sources in high-dose-rate brachytherapy using Geant4 Monte Carlo code},
  journal = {Journal of Contemporary Brachytherapy},
  year = {2023},
  volume = {15},
  number = {2},
  pages = {141--147},
  doi = {10.5114/jcb.2023.126238}
}
Kostenlos starten →