Criterion validity and accuracy of a heart rate monitor

Vinicius De Oliveira Damasceno, André Dos Santos Costa, Maria Clara Araújo Campello, Danilo Edson Souza, Reginaldo Gonçalves, Eduardo Zapaterra Campos, Tony Meireles Santos

Human Movement · 17.08.2021 · 23 (1) · 60-68

DOI: 10.5114/hm.2021.104188 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Human Movement, online erschienen am 17.08.2021. Verfasst von Vinicius De Oliveira Damasceno und 6 weiteren Autor:innen. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 41 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Heart rate (HR) monitors have recently started to use photoplethysmography, a technique which measures the light reflected by blood vessels and does not require the use of a chest strap. The aim of this study was to test the validity and accuracy of the Garmin® HR monitor, which measures HR at rest and during exercise utilizing the method of photoplethysmography.

Methods

The sample consisted of 28 males aged 18–32 years. Anthropometric measurements were collected and HR was concomitantly monitored with electrocardiography and with the Garmin® 735XT® device in 2 situations: at rest and during self-selected exercise. Descriptive statistics, linear regression, Bland-Altman plot, mean absolute error (MAE), and mean absolute percentage error (MAPE) were calculated for statistical analysis. Correlations between the HR measurement with electrocardiography and the Garmin® monitor at rest and during exercise were obtained (r = 0.93 and r = 0.96, respectively).

Results

The difference between Garmin® and electrocardiography HR values showed an error of –1.2 ± 3.3 bpm (rest), while the average error was positive at 0.7 ± 5.1 bpm. MAE and MAPE at rest equalled 2.2 ± 2.8 bpm and 3.3%, respectively. In addition, MAE and MAPE for exercise were 3.5 ± 3.8 bpm and 3.0%, respectively.

Conclusions

The Garmin Forerunner 735XT can be used at rest, as well as with walking and running activities of light, moderate, and vigorous intensities.

Zitieren

APA

Damasceno, V. D. O., Costa, A. D. S., Campello, M. C. A., Souza, D. E., Gonçalves, R., Campos, E. Z., & Santos, T. M. (2021). Criterion validity and accuracy of a heart rate monitor. Human Movement, 23(1), 60-68. https://doi.org/10.5114/hm.2021.104188

Vancouver

Damasceno VDO, Costa ADS, Campello MCA, Souza DE, Gonçalves R, Campos EZ, et al. Criterion validity and accuracy of a heart rate monitor. Human Movement. 2021;23(1):60-68. doi:10.5114/hm.2021.104188

BibTeX
@article{Damasceno2021,
  author = {Damasceno, Vinicius De Oliveira and Costa, André Dos Santos and Campello, Maria Clara Araújo and Souza, Danilo Edson and Gonçalves, Reginaldo and Campos, Eduardo Zapaterra and Santos, Tony Meireles},
  title = {Criterion validity and accuracy of a heart rate monitor},
  journal = {Human Movement},
  year = {2021},
  volume = {23},
  number = {1},
  pages = {60--68},
  doi = {10.5114/hm.2021.104188}
}
Kostenlos starten →