Radiotherapy of prostate and head-and-neck tumour: an optimal treatment planning comparison for intensity modulated radiotherapy and volumetric modulated arctherapy techniques
Journal of the Pakistan Medical Association · 2023
DOI: 10.47391/jpma.6447 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of the Pakistan Medical Association, 2023. Verfasst von Ghulam Murtaza. Beteiligte Einrichtung: Department of Medical Imaging, School of Health Sciences, University of Management and Technology, Lahore, Pakistan. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiation Oncology, Head and Neck Neoplasms. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
To evaluate and compare merits between intensity modulated radiotherapy and volumetric modulated arc-therapy radiotherapy techniques to determine which technique can achieve better treatment plan quality and efficient delivery.
Method
The retrospective study was conducted at the Radiation Oncology Department of SanBorotlo Hospital, Vicenza, Italy, in 2019, and comprised data from Jan 2019 to Dec 2019 related to prostate and head-and-neck patients in whom Pinnacle³ treatment planning system was used for optimisation with different prescribed doses and target geometries for intensity modulated radiotherapy and volumetric modulated arc-therapy techniques. Treatment plans were simulated using 6MV photon beam of SynergyS® Linac (Linear accelerator). The plan quality was evaluated using dose-volume indices for planning target-volume and organs-at-risk. ArcCHECK™ phantom was used for dose agreement verification between planed and delivered doses.
Results
Data of 8 patients was analysed. Intensity modulated radiotherapy and volumetric modulated arc-therapy treatment plan quality for prostate was found to be similar, but volumetric modulated arc-therapy had significant results for maximum dose (p=0.005). Intensity modulated radiotherapy and volumetric modulated arc-therapy plans for head-and-neck achieved adequate target coverage and sparing of organs at risk, and produced clinically acceptable treatment plans. The percentage of target coverage (p=0.001), dose maximum (p=0.013) and conformity index (p=0.000) were significant. A significant gain for all planning target volume dose-volume indices was noted (p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Imaging, School of Health Sciences, University of Management and Technology, Lahore, Pakistan.
Zitieren
APA
Murtaza, G. (2023). Radiotherapy of prostate and head-and-neck tumour: an optimal treatment planning comparison for intensity modulated radiotherapy and volumetric modulated arctherapy techniques. Journal of the Pakistan Medical Association. https://doi.org/10.47391/jpma.6447
Vancouver
Murtaza G. Radiotherapy of prostate and head-and-neck tumour: an optimal treatment planning comparison for intensity modulated radiotherapy and volumetric modulated arctherapy techniques. Journal of the Pakistan Medical Association. 2023. doi:10.47391/jpma.6447
BibTeX
@article{Murtaza2023,
author = {Murtaza, Ghulam},
title = {Radiotherapy of prostate and head-and-neck tumour: an optimal treatment planning comparison for intensity modulated radiotherapy and volumetric modulated arctherapy techniques},
journal = {Journal of the Pakistan Medical Association},
year = {2023},
doi = {10.47391/jpma.6447}
}
EUVIMED