Comparative Evaluation of TRIOS 5, Primescan, and Medit i900 Intraoral Scanners for Full-Arch Implant-Supported Prostheses: An in Vitro Study
Journal of Pioneering Medical Sciences · 05.11.2026 · 15 (10) · 70-81
DOI: 10.47310/jpms2026151007 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Pioneering Medical Sciences, online erschienen am 05.11.2026. Verfasst von Turki Almuraikhi und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Background
Accurate digital recording of implant positions is essential for the fabrication of full-arch implant-supported prostheses. The accuracy of intraoral scanners may vary because of differences in image acquisition and data-processing technologies.
Objective
To compare the trueness and precision of three intraoral scanners for full-arch implant-supported prostheses and evaluate their linear and angular deviations. Materials and Methods: A completely edentulous maxillary model containing six implant analogs was scanned using three intraoral scanners: TRIOS 5 (3Shape, Denmark), Primescan (Dentsply Sirona, Germany), and Medit i900 (Medit, Republic of Korea). Ten repeated full-arch scans were obtained with each scanner, resulting in 30 experimental datasets. A laboratory reference scan was used for three-dimensional comparison. Trueness and precision were evaluated using root mean square (RMS) deviations, while implant positional accuracy was assessed through linear and angular deviations. Statistical comparisons were performed using one-way analysis of variance or the Kruskal–Wallis test, as appropriate, with post-hoc pairwise comparisons adjusted for multiple testing. Statistical significance was set at P<0.05. Linear and angular deviations were measured at the six implant positions, and the mean deviation across the six positions was calculated for each scan. These scan-level values were used for comparisons among scanner systems.
Results
Mean trueness deviations were 42.6±7.8 µm for TRIOS 5, 58.4±10.1 µm for Primescan, and 76.9±13.5 µm for Medit i900. Corresponding precision deviations were 31.8±6.4 µm, 44.7±8.6 µm, and 61.2±11.3 µm, respectively. Mean linear deviations were 54.3±11.2 µm, 72.8±14.6 µm, and 96.5±18.7 µm, while angular deviations were 0.31±0.10°, 0.48±0.16°, and 0.71±0.22°, respectively. The manuscript reports overall differences among the scanner groups for all four outcomes (p< 0.001).
Conclusion
Under the standardized laboratory conditions of this in vitro study, TRIOS 5 demonstrated the lowest mean trueness, precision, linear, and angular deviations, followed by Primescan and Medit i900. These findings indicate differences in the measured accuracy of the three scanner systems within the experimental model. Further clinical investigations are required to determine whether these differences influence prosthesis fit and clinical outcomes.
Zitieren
APA
Almuraikhi, T., & Daghriri, A. (2026). Comparative Evaluation of TRIOS 5, Primescan, and Medit i900 Intraoral Scanners for Full-Arch Implant-Supported Prostheses: An in Vitro Study. Journal of Pioneering Medical Sciences, 15(10), 70-81. https://doi.org/10.47310/jpms2026151007
Vancouver
Almuraikhi T, Daghriri A. Comparative Evaluation of TRIOS 5, Primescan, and Medit i900 Intraoral Scanners for Full-Arch Implant-Supported Prostheses: An in Vitro Study. Journal of Pioneering Medical Sciences. 2026;15(10):70-81. doi:10.47310/jpms2026151007
BibTeX
@article{Almuraikhi2026,
author = {Almuraikhi, Turki and Daghriri, Alhussain},
title = {Comparative Evaluation of TRIOS 5, Primescan, and Medit i900 Intraoral Scanners for Full-Arch Implant-Supported Prostheses: An in Vitro Study},
journal = {Journal of Pioneering Medical Sciences},
year = {2026},
volume = {15},
number = {10},
pages = {70--81},
doi = {10.47310/jpms2026151007}
}