Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients
Ghana Medical Journal · 2025
DOI: 10.4314/gmj.v59i3.4 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Ghana Medical Journal, 2025. Verfasst von Charles Darko-Takyi und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Optometry and Vision Science, College of Health and Allied Sciences, University of Cape Coast, Ghana. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular, Fluorescein, Glaucoma. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
The interchangeability of Goldmann Applanation Tonometry (GAT) with or without fluorescein requires clarification among African populations. The present study aimed to determine the agreement between IOP measurement using GAT with and without fluorescein in consecutive patients in Ghana.
Design
A prospective cross-sectional clinic-based study design.
Setting
The study was conducted at the Ophthalmology and Optometry units of the UCC Eye Clinic, Cape Coast, Ghana.
Participants
A convenient sample of 104 consecutive patients who visited the clinic. The participants were randomised into two groups, where those in the first had their IOPs first measured without fluorescein and later with fluorescein on GAT; participants in the second group had their IOPs first measured with fluorescein and later without fluorescein after a 10-minute washout period. The central corneal thickness was measured using the Anterior Segment model of Zeiss Cirrus HD-OCT.
Results
There was no agreement between the unadjusted IOP measures with and without fluorescein with GAT using the Bland Altman analysis (p < 0.0001, Lower limit: -0.9433, Upper limit: 6.8914; 95% CI = -1.6097 to -.02769). There were no statistically significant differences in gender between unadjusted IOP measurements with (Mann-Whitney U test, p = 0.110) and without (Mann-Whitney U test, p = 0.083) fluorescein.
Conclusion
Unadjusted IOP measurement without fluorescein was approximately 3 mmHg significantly lower than with fluorescein. We recommend practitioners maintain the gold standard GAT with fluorescein to guide accurate glaucoma diagnosis and management decision-making.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Optometry and Vision Science, College of Health and Allied Sciences, University of Cape Coast, Ghana.
- University of Cape Coast Eye Clinic, Cape Coast, Ghana.
- Our Lady of Grace Catholic Mission Hospital, Bremen Esikuma, Ghana.
- Korle-Bu Teaching Hospital Eye Centre, Accra, Ghana.
Zitieren
APA
Darko-Takyi, C., Mathurin, Y. B., Essien, E., Owusu, S., Yaw, O. A., Osei, K. O., Holdbrook, S., & Owusu-Ansah, A. (2025). Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients. Ghana Medical Journal. https://doi.org/10.4314/gmj.v59i3.4
Vancouver
Darko-Takyi C, Mathurin YB, Essien E, Owusu S, Yaw OA, Osei KO, et al. Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients. Ghana Medical Journal. 2025. doi:10.4314/gmj.v59i3.4
BibTeX
@article{DarkoTakyi2025,
author = {Darko-Takyi, Charles and Mathurin, Youfegan B and Essien, Emmanuel and Owusu, Sandra and Yaw, Osei A and Osei, Kwame O and Holdbrook, Selina and Owusu-Ansah, Andrew},
title = {Agreement between intraocular pressure measurement using Goldmann applanation tonometry with and without fluorescein in consecutive Ghanaian patients},
journal = {Ghana Medical Journal},
year = {2025},
doi = {10.4314/gmj.v59i3.4}
}
EUVIMED