Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study
Korean Journal of Ophthalmology · 2025
DOI: 10.3341/kjo.2025.0016 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Korean Journal of Ophthalmology, 2025. Verfasst von Siamak Zarei-Ghanavati und 7 weiteren Autor:innen. Beteiligte Einrichtung: Eye Research Center, Mashhad University of Medical Sciences, Mashhad, Iran. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure, Keratoconus, Corneal Stroma. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Purpose
To compare intraocular pressure (IOP) readings from corneas with intracorneal corneal ring segments (ICRS) using various methods, including Goldmann applanation tonometry (GAT), Tonopen, corneal-compensated IOP from the Ocular Response Analyzer (ORA), and biomechanically corrected IOP from the Corneal Visualization Scheimpflug Technology (Corvis ST).
Methods
This cross-sectional observational study included participants who had undergone ICRS implantation with KeraRing at least 3 months before the study. The mean IOP recorded by different instruments was compared using analysis of variance. Agreement among the methods was assessed with Bland-Altman plots.
Results
A total of 54 eyes from 27 participants were enrolled. The mean IOP measured by Tonopen was significantly lower in the center compared to the peripheral quadrants (p < 0.001). IOP measured by GAT was significantly lower than that measured by Tonopen (13.02 ± 2.31 mmHg vs. 14.50 ± 2.91 mmHg, p = 0.021). There were no significant differences between the IOP measurements provided by Tonopen, ORA, and Corvis ST. The corneal-compensated IOP from ORA and biomechanically corrected IOP from Corvis ST had the highest correlation, with a weak intraclass correlation coefficient of 0.38.
Conclusions
IOP measurements using Tonopen were significantly lower in the central 5-mm zone compared to other quadrants. GAT measurements were significantly lower than those from Tonopen. Different measurement tools did not show a strong correlation. Corvis ST (biomechanically corrected IOP) tended to present lower readings at higher IOP levels in eyes with ICRS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Eye Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
- Glaucoma Service, Farabi Eye Hospital, Tehran University of Medical Sciences, Tehran, Iran.
- Eye Research Center, Amiralmomenin Hospital, Guilan University of Medical Sciences, Rasht, Iran.
Zitieren
APA
Zarei-Ghanavati, S., Tabatabaei, S. M., Gholamhoseinpour-Omran, S., Hosseinikhah-Manshadi, H., Banan, S., Aminizade, M., Esmaili, K., & Azaripour, E. (2025). Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study. Korean Journal of Ophthalmology. https://doi.org/10.3341/kjo.2025.0016
Vancouver
Zarei-Ghanavati S, Tabatabaei SM, Gholamhoseinpour-Omran S, Hosseinikhah-Manshadi H, Banan S, Aminizade M, et al. Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study. Korean Journal of Ophthalmology. 2025. doi:10.3341/kjo.2025.0016
BibTeX
@article{ZareiGhanavati2025,
author = {Zarei-Ghanavati, Siamak and Tabatabaei, Seyed Mehdi and Gholamhoseinpour-Omran, Samaneh and Hosseinikhah-Manshadi, Hamed and Banan, Saeed and Aminizade, Mehdi and Esmaili, Kosar and Azaripour, Ebrahim},
title = {Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study},
journal = {Korean Journal of Ophthalmology},
year = {2025},
doi = {10.3341/kjo.2025.0016}
}
EUVIMED