Lack of consensus in inter-laboratory haematology results in selected laboratories in the southern and northern zones of Ghana
Ghana Medical Journal · 2023
DOI: 10.4314/gmj.v57i3.8 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ghana Medical Journal, 2023. Verfasst von Ibrahim B Halidu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Laboratory Science, School of Allied Health Sciences, University of Cape Coast, Ghana. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 2. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
To assess the inter-laboratory comparability and intra-assay reproducibility of full blood count (FBC) results.
Design
Exploratory cross-sectional study.
Setting
Three and two selected medical laboratories in the northern and southern zones, respectively.
Participants
Forty-nine individuals per zone; 16 type 2 diabetes mellitus, 16 with HbAS haemoglobin type and 17 normal samplesIntervention: Each sample was run eleven times through the analysers in the participating laboratories to evaluate intra-laboratory reproducibility and comparability of FBC results.Main Outcome Measure: Intra-laboratory reproducibility was evaluated using %coefficient variation (%CV). Inter-laboratory comparisons were assessed through t-test or One-Way ANOVA for two-sample and three-sample tests. All statistical testing was undertaken using the two-tailed assumption.
Result
Statistically significantly different haemoglobin levels were estimated in both northern and southern zones (mean difference 0.00 g/dL to 3.75 g/dL vs 0.18 g/dL to 1.92 g/dL respectively). Also, total WBC counts significantly differed across laboratories in both northern and southern zones (mean difference 0.15 x109/L - 3.86 x109/L vs 0.02 x109/L to 1.39 x109/L respectively). Furthermore, platelet counts significantly differed across the participating laboratories in the northern and southern zones (mean difference 0.40 x109/L to 299.76 x109/L vs 5.7 x109/L to 76.9 x109/L respectively). Moreover, there was evidence of non-reproducibility of results within the respective laboratories in each zone as the respective %CV were outside the acceptable limits.
Conclusion
The intra-laboratory non-reproducibility and inter-laboratory non-comparability of FBC results highlight the need to establish a national quality assessment scheme to harmonise laboratory practices nationwide.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Medical Laboratory Science, School of Allied Health Sciences, University of Cape Coast, Ghana.
- Department of Biomedical Sciences, School of Allied Health Sciences, University of Cape Coast, Ghana.
- Medical OPD, Cape Coast Teaching Hospital, Cape Coast, Ghana.
Zitieren
APA
Halidu, I. B., Gafa, A. X., Blanney, S. D. K., Barimah, B. T., Akan-Enge, D., Boachie, J., Kontor, K. A., & Adu, P. (2023). Lack of consensus in inter-laboratory haematology results in selected laboratories in the southern and northern zones of Ghana. Ghana Medical Journal. https://doi.org/10.4314/gmj.v57i3.8
Vancouver
Halidu IB, Gafa AX, Blanney SDK, Barimah BT, Akan-Enge D, Boachie J, et al. Lack of consensus in inter-laboratory haematology results in selected laboratories in the southern and northern zones of Ghana. Ghana Medical Journal. 2023. doi:10.4314/gmj.v57i3.8
BibTeX
@article{Halidu2023,
author = {Halidu, Ibrahim B and Gafa, Amos X and Blanney, Samuel D K and Barimah, Benjamin T and Akan-Enge, David and Boachie, Joseph and Kontor, Kate A and Adu, Patrick},
title = {Lack of consensus in inter-laboratory haematology results in selected laboratories in the southern and northern zones of Ghana},
journal = {Ghana Medical Journal},
year = {2023},
doi = {10.4314/gmj.v57i3.8}
}
EUVIMED