Evaluation of Different Methods of Haemoglobin Estimation Among Malaria‐Infected Patients

Mina Ofosu iD, Jessica Ntiamoah, Patricia Sam Barko iD, Prince Elikem Torli iD

Journal of Clinical Laboratory Analysis · 2026

DOI: 10.1002/jcla.70377 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Laboratory Analysis, 2026. Verfasst von Mina Ofosu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Technology, Faculty of Health Sciences, Kumasi Technical University, Kumasi, Ghana. Hauptthemen laut MEDLINE: Hemoglobins, Malaria, Hemoglobinometry. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Haemoglobin (Hb) estimation is essential for assessing anaemia, particularly in malaria‐endemic settings. Automated Full Blood Count (FBC) analysers, cyanmethemoglobin spectrophotometry, and portable Hb meters are commonly used, but differences in analytical performance may affect measurements.

Methods

This cross‐sectional study compared Hb measurements obtained using an automated FBC analyser, the cyanmethemoglobin method, and a portable Hb meter among 100 malaria‐infected patients aged 16–65 years at Kumasi South Hospital, Ghana. Measurements were compared using Pearson's correlation coefficient, intraclass correlation coefficient (ICC), Lin's concordance correlation coefficient (CCC), Bland–Altman analysis, percentage error, and Passing–Bablok regression.

Results

The three methods showed very strong correlations ( r = 0.955–0.992, p < 0.001), excellent ICC (single measures = 0.969; average measures = 0.990), and high concordance (CCC = 0.95–0.99). Mean biases were −0.06 g/dL for FBC analyser versus portable Hb meter, −0.25 g/dL for FBC analyser versus cyanmethemoglobin, and −0.20 g/dL for portable Hb meter versus cyanmethemoglobin. The corresponding 95% limits of agreement were −0.59 to 0.47, −1.48 to 0.98, and −1.35 to 0.95 g/dL, respectively. Passing–Bablok slopes were 1.00 for all comparisons. The FBC analyser and portable Hb meter showed the closest agreement.

Conclusion

The three methods demonstrated good agreement and broadly comparable Hb measurements. The findings support the practical use of portable Hb meters where point‐of‐care testing is required, while recognising that agreement does not establish diagnostic accuracy or analytical validation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemoglobinsMalariaHemoglobinometryHumansAdultMiddle AgedFemaleAdolescentMaleYoung AdultCross-Sectional StudiesAgedGhanaBlood Cell CountAnemiaReproducibility of Results

Stichwörter der Autor:innen

agreementcyanmethemoglobinfull blood count analyserhaemoglobin estimationhaemoglobin metermalaria

Substanzen

Hemoglobins

Autor:innen und Institutionen

  • Department of Laboratory Technology, Faculty of Health Sciences, Kumasi Technical University, Kumasi, Ghana.

ORCID-Profile: Mina Ofosu, Patricia Sam Barko, Prince Elikem Torli

Zitieren

APA

Ofosu, M., Ntiamoah, J., Sam Barko, P., & Torli, P. E. (2026). Evaluation of Different Methods of Haemoglobin Estimation Among Malaria‐Infected Patients. Journal of Clinical Laboratory Analysis. https://doi.org/10.1002/jcla.70377

Vancouver

Ofosu M, Ntiamoah J, Sam Barko P, Torli PE. Evaluation of Different Methods of Haemoglobin Estimation Among Malaria‐Infected Patients. Journal of Clinical Laboratory Analysis. 2026. doi:10.1002/jcla.70377

BibTeX
@article{Ofosu2026,
  author = {Ofosu, Mina and Ntiamoah, Jessica and Sam Barko, Patricia and Torli, Prince Elikem},
  title = {Evaluation of Different Methods of Haemoglobin Estimation Among Malaria‐Infected Patients},
  journal = {Journal of Clinical Laboratory Analysis},
  year = {2026},
  doi = {10.1002/jcla.70377}
}
Kostenlos starten →