Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators
Respiratory Care · 2023
DOI: 10.4187/respcare.10989 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiratory Care, 2023. Verfasst von Lea Bergez und 2 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pediatrics and Neonatal Critical Care, "A.Beclere" Medical Center, Paris Saclay University Hospitals, APHP, Paris, France. Hauptthemen laut MEDLINE: Respiration, Artificial, High-Frequency Ventilation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
BACKGROUND: Insufficient data are available about the noise produced by modern neonatal ventilators. We aimed to measure their noise under different ventilatory modes and parameters. METHODS: This was a bench study measuring the noise produced by 9 neonatal ventilators set in conventional or high-frequency oscillatory ventilation (HFOV), nasal mask-delivered CPAP with variable- or continuous-flow configuration, or bi-level positive airway pressure (considered as noninvasive ventilation [NIV]). Conventional ventilation and HFOV were tested in 2 distinct settings with moderate or higher parameters. Sound measurements were performed inside and outside an incubator mimicking the clinical setting and using a high-end meter meeting the international ISO 226:2003 standard. RESULTS: Four ventilators remained below the internationally recommended safety threshold but only for measurements outside the incubator. Conventional ventilation (49.1 [3.4] dBA) and HFOV (56.3 [5.2] dBA) were the least and most noisy respiratory support technique, respectively. Noise was greater inside than outside the incubators (P < .0001) and different between the ventilators (P < .0001); better results were achieved by Servo-u and Fabian family devices for conventional ventilation; by fabian HFO for HFOV; and by Servo-u, VN500, and fabian family devices for CPAP and NIV. Noise levels were similar when using moderate or higher parameters in conventional ventilation (P = .81) and in HFOV (P = .45). CONCLUSIONS: Modern ventilators often produce relevant noise, independent of the respiratory support modality, with acceptable noise levels being measured only outside the incubator. Better results were achieved with Servo-u, VN500, and Fabian family devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Pediatrics and Neonatal Critical Care, "A.Beclere" Medical Center, Paris Saclay University Hospitals, APHP, Paris, France.
- Division of Pediatrics and Neonatal Critical Care, "A.Beclere" Medical Center, Paris Saclay University Hospitals, APHP, Paris, France; and Physiopathology and Therapeutic Innovation Unit-INSERM U999, Paris Saclay Univer…
Zitieren
APA
Bergez, L., Jourdain, G., & De Luca, D. (2023). Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators. Respiratory Care. https://doi.org/10.4187/respcare.10989
Vancouver
Bergez L, Jourdain G, De Luca D. Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators. Respiratory Care. 2023. doi:10.4187/respcare.10989
BibTeX
@article{Bergez2023,
author = {Bergez, Lea and Jourdain, Gilles and De Luca, Daniele},
title = {Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators},
journal = {Respiratory Care},
year = {2023},
doi = {10.4187/respcare.10989}
}
EUVIMED