Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery
Journal of Medical Physics · 2021-10 · 46 (4) · 286-294
DOI: 10.4103/jmp.jmp_18_21 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Physics, 2021-10. Verfasst von Soumya Roy und 2 weiteren Autor:innen. Beteiligte Einrichtung: Depertment of Radiation Oncology, Manipal Hospitals, Dwarka, New Delhi, India. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Aim
To determine the multileaf collimator positional error (MLC-PE) during volumetric modulated arc therapy (VMAT) delivery by studying the time-dependent MLC velocity in mathematically derivable trajectories such as straight line and conic sections. Materials and Methods: VMAT delivery is planned in a way that MLCs are moving in a locus which can be defined by mathematical functions such as linear, parabolic, or circular velocity (PV or CV). The VMAT delivery was interrupted either once or multiple times during the delivery and projection images of the same were acquired in electronic portal imaging device. MLC-PE was then analyzed as a function of dose rate (DR), and MLC speed (SP) and number of interruptions in treatment delivery. In VMAT delivery with linear MLC motion, the delivery was interrupted either once (linear motion single interruption) or multiple (three) times (linear motion multiple interruptions). For PV and CV MLC velocity, the MLC motions are interrupted multiple times.
Results
The maximum individual error obtained (DR of 35 MU/min, SP of 2.0 cm/s) was 1.96 ± 0.1 mm. Only 4.4% of MLCs showed ≥ ±1 mm positional error. When the treatment delivery is interrupted multiple times in VMAT delivery, the influence of interruption in MLC-PE overwhelmed the influence by DR and SP. For a sub-group analysis of independent and dependent variables, the mean MLC-PE was 0.18 ± 0.4 mm 0.19 ± 0.42 mm, respectively.
Conclusion
Determination of MLC-PEs using a mathematical function without approximation indicates that MLC-PE is not a function of MLC speed. In less than 5% of the studied scenarios, the MLC-PE exceeds its tolerance value (±1 mm). The MLC-PE is significantly less in modern machines due to advancements in the delivery mechanism.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Depertment of Radiation Oncology, Manipal Hospitals, Dwarka, New Delhi, India.
- Department of Physics, GLA University, Mathura, Uttar Pradesh, India.
- Department of Radiotherapy, Apollo Multispeciality Hospitals, Kolkata, West Bengal, India.
Zitieren
APA
Roy, S., Sarkar, B., & Pradhan, A. (2021). Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery. Journal of Medical Physics, 46(4), 286-294. https://doi.org/10.4103/jmp.jmp_18_21
Vancouver
Roy S, Sarkar B, Pradhan A. Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery. Journal of Medical Physics. 2021;46(4):286-294. doi:10.4103/jmp.jmp_18_21
BibTeX
@article{Roy2021,
author = {Roy, Soumya and Sarkar, Biplab and Pradhan, Anirudh},
title = {Determination of Multileaf Collimator Positional Errors as a Function of Dose Rate, Speed, and Delivery Interruption for Volumetric-Modulated Arc Therapy Delivery},
journal = {Journal of Medical Physics},
year = {2021},
volume = {46},
number = {4},
pages = {286--294},
doi = {10.4103/jmp.jmp_18_21}
}
EUVIMED